DOI 10.1097/mao.0000000000000604
Effects of body tilt on multifrequency admittance tympanometry.
Аннотация
INTRODUCTION
Hydrops and abnormalities of inner fluid pressure are involved in some otologic diseases such as Ménière's disease (MD). However, demonstrating abnormal perilymphatic or endolymphatic pressure is challenging. Multifrequency tympanometry studies in MD patients demonstrated an increase of the width of conductance tympanograms (outside an attack) compared with controls. To confirm that the increase in conductance width is caused by hyperpressure and not hypopressure in these patients tested outside an attack, we assessed the effect of changes in inner ear fluid pressure caused by body tilt on the results of multifrequency admittancemetry tympanograms.
MATERIALS AND
METHODS
A multifrequency tympanometry including conductance (G) tympanogram at 2 kHz and resonance frequency measurements were performed in 20 volunteers (40 ears) free of otologic or neurologic disease. The measures were collected in three different positions: vertical, supine, and Trendelenburg positions.
RESULTS
Changes in inner ear fluid pressure caused by body tilt induced an increase in the width of G tympanograms. In the vertical position, the mean value was 141.7 ± 56.5 daPa; in the supine position, it increased to 158 ± 58.3 daPa; and increased even more in the Trendelenburg position (20 degrees), with a mean of 184 ± 69.6 daPa (p < 0.01). Resonance frequency also increased in the Trendelenburg position.We conclude that the increased width of G tympanograms in MD patients outside an attack may be caused by an increase in inner ear fluid pressure.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.