OBJECTIVE: A cerebrospinal fluid leak is one of the most serious complications in otolaryngology. It may occur as a result of injury to the skull base, typically traumatic or iatrogenic. While the presence of a leak is often discerned in the emergent setting, distinguishing normal secretions from those containing cerebrospinal fluid can be difficult during postoperative visits in the clinic. As most current laboratory-based assays are labor intensive and require several days to result, we aim to develop a more user-friendly and rapid point-of-care cerebrospinal fluid detection device.
STUDY DESIGN: Our laboratory developed a barcode-style lateral-flow immunoassay utilizing antibodies for beta-trace protein, a protein abundant in and specific for cerebrospinal fluid, with a concentration of 1.3 mg/L delineating a positive result.
SETTING: Tertiary medical center.
SUBJECTS AND METHODS: Tests with known concentrations of resuspended beta-trace protein and the contents of discarded lumbar drains (presumed to contain cerebrospinal fluid) were performed to validate our novel device.
RESULTS: Our results demonstrate the ability of our device to semiquantitatively identify concentrations of beta-trace protein from 0.3-90 mg/L, which is within the required range to diagnose a leak, thus making beta-trace protein an excellent target for rapid clinical detection.
CONCLUSION: Herein we detail the creation and initial validation of the first point-of-care cerebrospinal fluid detection device. This device is a feasible method to more efficiently and cost-effectively identify cerebrospinal fluid leaks, minimize costs, and improve patient outcomes.
STUDY DESIGN: Our laboratory developed a barcode-style lateral-flow immunoassay utilizing antibodies for beta-trace protein, a protein abundant in and specific for cerebrospinal fluid, with a concentration of 1.3 mg/L delineating a positive result.