Eyewatch adjustable drainage device in vitrectomized eyes with refractory glaucoma.
Аннотация
BACKGROUND
To describe the efficacy and safety of the adjustable system eyeWatch in vitrectomized glacomatous eyes.
METHODS
Prospective, non-comparative, small case series of 2 patients who underwent glaucoma drainage device implant with the eyeWatch system. Intraocular pressure, number of medications and early and late complications were evaluated.
RESULTS
None of the operated eyes developed early or late complications. After 1 week post operatively, the IOP raised at 28 mmHg in the first eye and 25 mmHg in the second one. The eyeWatch Pen was used to open and set the device into position 3 (that means implant half opened) After waiting for 30 min, IOP was 15 mmHg and 11 mmHg, respectively. Thirty days after surgery we set the EyeWatch into position 0 (that means implant full opened) even though the IOP was under control. A new IOP measurement was done and we registered that IOP was 14 mmHg and 10 mmHg respectively, and it remained stable after six months of follow up.
CONCLUSION
In conclusion, in our case series the use of eyeWatch adjustable system allows a safe and gradual regulation of the flow during post-operative management also in vitrectomized eyes. Considering the percentage of hypotony related complications in vitrectomized eyes this system is an effective and safe technique to handle glaucoma in post vitreoretinal surgery eyes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.