MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/ijg.0000000000002231

Медицинская библиотека · Journal of glaucoma · 2023

Negative Pressure Application via a Multi-Pressure Dial to Lower IOP in Patients with Suspected Glaucoma or Open Angle Glaucoma.

Samuelson Thomas W, Ferguson Tanner J, Brubaker Jacob W, Bacharach Jason, Gallardo Mark J, Chaglasian Michael, Comer George, Yoo Paul

Рандомизированное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PRÉCIS: The multi-pressure dial applies localized periocular negative pressure to safely and effectively lower IOP and represents the first non-invasive, non-pharmacologic device for IOP reduction.

OBJECTIVE

To evaluate the safety and effectiveness of the Multi-Pressure Dial (MPD) system, a device that applies periocular negative pressure to lower intraocular pressure (IOP).

SETTING

6 investigational sites, United States.

DESIGN

Prospective, assessor-masked, randomized controlled trial.

METHODS

Subjects with suspected glaucoma, ocular hypertension (OHTN), and open angle glaucoma (OAG) with baseline IOP ≥13 mmHg and ≤32 mmHg were enrolled. One eye of each subject was randomized to receive negative pressure application; the fellow eye served as a control. The study eye negative pressure setting was programmed for 60% of the baseline IOP. The primary effectiveness endpoint was the proportion of study eyes versus control eyes achieving an IOP reduction ≥20% at Day 90. Secondary endpoints included the proportion of eyes achieving an IOP reduction ≥25% at Day 90 as well as the proportion of eyes achieving an IOP reduction ≥20% at Days 30 and 60.

RESULTS

116 eyes of 58 subjects completed the study. At the Day 90 visit, 89.7% ( n =52) of study eyes versus 3.4% ( n =2) of control eyes achieved an IOP reduction ≥20% ( P <0.001). At Day 90, 77.6% ( n =45) of study eyes achieved a ≥25% IOP reduction compared to 1.7% ( n =1) of control eyes ( P <0.001). The most commonly reported adverse events were lid (17.2% study eye, 7.8% control eye) and periorbital edema (14.1% study eye, 10.9% control eye).

CONCLUSIONS

This trial demonstrates that the MPD safely and effectively lowers IOP in a group of patients that included glaucoma suspects, OHTN, and patients with OAG.

Для цитирования: Samuelson Thomas W, Ferguson Tanner J, Brubaker Jacob W, Bacharach Jason, Gallardo Mark J, Chaglasian Michael, Comer George, Yoo Paul. Negative Pressure Application via a Multi-Pressure Dial to Lower IOP in Patients with Suspected Glaucoma or Open Angle Glaucoma.. Journal of glaucoma. 2023. DOI: 10.1097/ijg.0000000000002231.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.