Validation of xMAP SARS-CoV-2 Multi-Antigen IgG assay in Nigeria.
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Iriemenam Nnaemeka C, Ige Fehintola A, Greby Stacie M, Mpamugo Augustine, Abubakar Ado G, Dawurung Ayuba B, Esiekpe Mudiaga K, Thomas Andrew N, Okoli Mary U, Awala Samuel S, Ugboaja Blessing N, Achugbu Chimaoge C, Odoh Ifeanyichukwu, Nwatu Felicia D, Olaleye Temitope, Akayi Loveth, Akinmulero Oluwaseun O, Dattijo Joseph, Onokevbagbe Edewede, Okunoye Olumide, Mba Nwando, Agala Ndidi P, Uwandu Mabel, Aniedobe Maureen, Stafford Kristen A, Abimiku Alash'le, Hamada Yohhei, Swaminathan Mahesh, Okoye McPaul I, Steinhardt Laura C, Audu Rosemary
Аннотация
METHODS: Validation of the xMAP SARS-CoV-2 Multi-Antigen IgG assay was carried out using well-characterized SARS-CoV-2 reverse transcription polymerase chain reactive positive (97) and pre-COVID-19 pandemic (86) plasma panels. Cross-reactivity was assessed using pre-COVID-19 pandemic plasma specimens (213) from the 2018 Nigeria HIV/AIDS Indicator and Impact Survey (NAIIS).
RESULTS: The overall sensitivity of the xMAP SARS-CoV-2 Multi-Antigen IgG assay was 75.3% [95% CI: 65.8%- 82.8%] and specificity was 99.0% [95% CI: 96.8%- 99.7%]. The sensitivity estimate increased to 83.3% [95% CI: 70.4%- 91.3%] for specimens >14 days post-confirmation of diagnosis. However, using the NAIIS pre-pandemic specimens, the false positivity rate was 1.4% (3/213).
CONCLUSIONS: Our results showed overall lower sensitivity and a comparable specificity with the manufacturer's validation. There appears to be less cross-reactivity with NAIIS pre-pandemic COVID-19 specimens using the xMAP SARS-CoV-2 Multi-Antigen IgG assay. In-country SARS-CoV-2 serology assay validation can help guide the best choice of assays in Africa.