Seidman Michael D, Vivek Prashant, Dickinson William
Аннотация
OBJECTIVE: The purpose of the study was to analyze changes in neural response telemetry using the Cochlear Nucleus 24 Contour before and after stylet removal in a human model.
STUDY DESIGN: Prospective study.
SETTING: Tertiary referral center.
PATIENTS: Twelve patients (23-72 years old) undergoing cochlear implantation, using the Nucleus 24 Contour implant, secondary to congenital and age-related hearing loss.
INTERVENTION: Cochlear implantation with Cochlear Nucleus 24 Contour implant.
MAIN OUTCOME MEASURE: Neural response telemetry thresholds were recorded initially with the stylet in and then with the stylet out during the implant procedure.
RESULTS: Stimulus levels to obtain neural response telemetry threshold after stylet removal were statistically lower after the stylet was removed.
CONCLUSION: The Nucleus 24 Contour achieves a perimodiolar position once the stylet is removed, and this leads to reduction in the current required to elicit a threshold neural response telemetry response.
STUDY DESIGN: Prospective study.
SETTING: Tertiary referral center.
PATIENTS: Twelve patients (23-72 years old) undergoing cochlear implantation, using the Nucleus 24 Contour implant, secondary to congenital and age-related hearing loss.
INTERVENTION: Cochlear implantation with Cochlear Nucleus 24 Contour implant.
MAIN OUTCOME MEASURE: Neural response telemetry thresholds were recorded initially with the stylet in and then with the stylet out during the implant procedure.
RESULTS: Stimulus levels to obtain neural response telemetry threshold after stylet removal were statistically lower after the stylet was removed.