Early outcomes with a fully retrievable SinoCrown transcatheter heart valve in patients with severe aortic stenosis.
Аннотация
BACKGROUND
This study summarizes and analyzes data from patients suffering from symptomatic aortic stenosis who successfully underwent transcatheter aortic valve implantation (TAVI) using a novel, completely retrievable transcatheter heart valve.
METHODS
We included patients who underwent a TAVI procedure with SinoCrown valves at our center between December 2021 and September 2022. We collected 1-year follow-up data on survival, complications, echocardiographic results, New York Heart Association functional class in heart failure, and patient-reported health-related quality of life outcomes.
RESULTS
Eight successive patients (73.3 ± 4.3 years) were included in the study, with a median Society of Thoracic Surgery risk score of 4.26%. The procedure had a 100% success rate. Median postoperative discharge time was 7 days, with no 30-day hospital readmissions. Postoperative aortic valve hemodynamics improved, indicated by decreased transvalvular flow velocity compared with preoperative values (1.9 ± 0.2 vs. 4.9 ± 0.2 m/s, p < 0.0001). The median and maximum follow-up times were 8 and 12 months, respectively. During the follow-up period, there were no serious complications such as death, stroke, valve embolization, or high-grade atrioventricular block.
CONCLUSIONS
The results from eight initial TAVI cases performed with the SinoCrown valve demonstrated promising safety and efficacy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.