DOI 10.1093/ehjimp/qyag087

Медицинская библиотека · European heart journal. Imaging methods and practice · 2026

Fully automated deep learning MAPSE: retrospective analysis and real-time clinical application.

Haga Maria Muan, Katla Nora Lindeman, Holmstrøm Vegard, Holte Espen, Stølen Stian, Stensæth Knut Haakon, Østvik Andreas, Løvstakken Lasse, Dalen Håvard, Smistad Erik, Grenne Bjørnar

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Mitral annular plane systolic excursion (MAPSE) is an accessible echocardiographic measure of left ventricular (LV) function. However, manual measurement methods are operator-dependent and time-consuming. We developed a multistep deep learning (DL) method for off-line and real-time fully automated MAPSE estimation, and aimed to assess agreement, reproducibility, time efficiency, and feasibility compared with standard manual measurements.

METHODS AND

RESULTS

The DL-based method was evaluated in two retrospective cohorts (n = 1775) and one prospective cohort (n = 51). Agreement between DL-MAPSE on B-mode images and experts' manual M-mode measurements was evaluated in all datasets. Evaluation of test-retest reproducibility, time efficiency using real-time analysis during acquisition, and agreement with cardiac magnetic resonance (CMR)-imaging were performed in subsets of the datasets. DL-MAPSE demonstrated good agreement with manual measurements, with bias 2.9 mm (95% CI 2.8-3.0 mm) and Pearson coefficient 0.81 (95% CI 0.79-0.84) in the primary dataset, and a lower bias of 1.0 mm against CMR-MAPSE compared with -2.1 mm using manual M-mode. Both DL and manual measurements showed good test-retest reproducibility (ICC 0.82 and 0.76, respectively). Real-time DL measurements reduced measurement and acquisition time by 51% (mean 1 min 50 s) per examination. The DL method demonstrated excellent feasibility (96%).

CONCLUSION

This novel DL method for fully automated MAPSE demonstrated excellent feasibility, robust reproducibility, and good agreement with both manual M-mode and CMR-derived measurements. Automated DL-MAPSE could substantially reduce analysis time and enhance reproducibility, increasing its clinical value as a marker of LV systolic function.

Для цитирования: Haga Maria Muan, Katla Nora Lindeman, Holmstrøm Vegard, Holte Espen, Stølen Stian, Stensæth Knut Haakon, Østvik Andreas, Løvstakken Lasse, Dalen Håvard, Smistad Erik, Grenne Bjørnar. Fully automated deep learning MAPSE: retrospective analysis and real-time clinical application.. European heart journal. Imaging methods and practice. 2026. DOI: 10.1093/ehjimp/qyag087.