Artificial organs · 2025

Hydraulic In Vitro Characterization of MR-Conditional Blood Pumps.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Schulte Dominik T, Renggli Marcel, Richter Henning, Del Chicca Francesca, Hofmann Michael, Schmiady Martin O, Schmid Daners Marianne

Аннотация

BACKGROUND: Many cardiac surgical procedures use cardiopulmonary bypass (CPB), which in neonates requires the heart-lung machine (HLM) to be positioned close to the patient due to their small circulating blood volume. The absence of an MR-compatible blood pump to be used in CPB remains a key challenge for studying perioperative brain injury mechanisms in this high-risk group. This study aims to take the first step toward MR-conditional HLMs by developing hardware, verifying pump hydraulics, and ensuring MR compatibility.

METHODS: This study presents three MR-conditional blood pump prototypes: a roller pump, a non-occlusive roller pump, and a centrifugal pump. MR compatibility was assessed by monitoring for imaging interference during scanning. Hydraulic performance was evaluated with pressure-flow diagrams using a mock circulation test bench.

RESULTS: None of the prototypes interfered with MR imaging, and although SNR was reduced by -8.43% ± 7.96%, image quality remained sufficient for reliable assessment of relevant brain regions. Roller and non-occlusive pumps maintained stable flow across pressure heads, with reductions of -4.46 ± 8.02 and -119.33 ± 18.22 mL/min, respectively. The centrifugal pump exhibited pressure-dependent performance (slope -2.21 ± 0.45 mmHg/[L/min]). All pumps generated non-pulsatile flow (SHE < 1000 erg/cm3).

CONCLUSION: All three pumps meet basic MR-conditionality and flow requirements, supporting their potential for use in MRI-guided studies during neonatal cardiac surgery.