DOI 10.1111/aor.70041

Медицинская библиотека · Artificial organs · 2025

Hydraulic In Vitro Characterization of MR-Conditional Blood Pumps.

Schulte Dominik T, Renggli Marcel, Richter Henning, Del Chicca Francesca, Hofmann Michael, Schmiady Martin O, Schmid Daners Marianne

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Many cardiac surgical procedures use cardiopulmonary bypass (CPB), which in neonates requires the heart-lung machine (HLM) to be positioned close to the patient due to their small circulating blood volume. The absence of an MR-compatible blood pump to be used in CPB remains a key challenge for studying perioperative brain injury mechanisms in this high-risk group. This study aims to take the first step toward MR-conditional HLMs by developing hardware, verifying pump hydraulics, and ensuring MR compatibility.

METHODS

This study presents three MR-conditional blood pump prototypes: a roller pump, a non-occlusive roller pump, and a centrifugal pump. MR compatibility was assessed by monitoring for imaging interference during scanning. Hydraulic performance was evaluated with pressure-flow diagrams using a mock circulation test bench.

RESULTS

None of the prototypes interfered with MR imaging, and although SNR was reduced by -8.43% ± 7.96%, image quality remained sufficient for reliable assessment of relevant brain regions. Roller and non-occlusive pumps maintained stable flow across pressure heads, with reductions of -4.46 ± 8.02 and -119.33 ± 18.22 mL/min, respectively. The centrifugal pump exhibited pressure-dependent performance (slope -2.21 ± 0.45 mmHg/[L/min]). All pumps generated non-pulsatile flow (SHE < 1000 erg/cm3).

CONCLUSION

All three pumps meet basic MR-conditionality and flow requirements, supporting their potential for use in MRI-guided studies during neonatal cardiac surgery.

Для цитирования: Schulte Dominik T, Renggli Marcel, Richter Henning, Del Chicca Francesca, Hofmann Michael, Schmiady Martin O, Schmid Daners Marianne. Hydraulic In Vitro Characterization of MR-Conditional Blood Pumps.. Artificial organs. 2025. DOI: 10.1111/aor.70041.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.