Comparison of vestibular schwannoma visualization between 0.55 T and 1.5 T MRI.

Wiesmueller Marco, Kopp Markus, Sievert Matti, May Matthias S, Nagel Armin M, Iro Heinrich, Uder Michael, Heiss Rafael

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Vestibular schwannoma (VS) is the most common mass of the internal auditory canal (IAC) and is responsible for unilateral sensorineural hearing loss. Magnetic resonance imaging (MRI) at 1.5 T and 3 T is the standard of care for the evaluation of VS, and the feasibility of using modern low-field MRI for imaging of the IAC has not yet been elucidated. Hence, the purpose of this prospective study was to assess image quality and diagnostic performance of a modern 0.55 T MRI.

MATERIALS AND

METHODS

Fifty-six patients with known unilateral VS underwent routine MRI of the IAC at 1.5 T, followed immediately by 0.55 T MRI. Two radiologists independently evaluated the image quality, conspicuity of VS, diagnostic confidence, and image artifacts separately for isotropic T2-weighted SPACE images and for transversal and coronal T1-weighted fat-saturated contrast-enhanced images at 1.5 T and 0.55 T using 5-point Likert scales. In a second independent reading, both readers assessed lesion conspicuity and subjective diagnostic confidence in a direct comparison of 1.5 T and 0.55 T images.

RESULTS

Image quality of transversal T1-weighted images (p = 0.13 and p = 0.16 for Reader 1 and Reader 2, respectively) and T2-weighted SPACE images (p = 0.39 and p = 0.58) were rated equally at 1.5 T and 0.55 T by both readers, whereas image quality of coronal T1-weighted images was superior at 1.5 T (p = 0.009 and p = 0.001). Analysis of the conspicuity of VS, diagnostic confidence, and image artifacts of all sequences revealed no significant differences between 1.5 T and 0.55 T. In the direct comparison of 1.5 T with 0.55 T images, there were no significant differences in lesion conspicuity or diagnostic confidence for any sequence (p = 0.60-0.73).

CONCLUSIONS

Modern low-field MRI at 0.55 T provided a sufficient diagnostic image quality and seems feasible for the evaluation of VS of the IAC.

Для цитирования: Wiesmueller Marco, Kopp Markus, Sievert Matti, May Matthias S, Nagel Armin M, Iro Heinrich, Uder Michael, Heiss Rafael. Comparison of vestibular schwannoma visualization between 0.55 T and 1.5 T MRI.. European journal of radiology. 2023. DOI: 10.1016/j.ejrad.2023.110927.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.