[Tensometry - development of a measurement technique of the tube function in a temporal bone model].

Аннотация на английскомСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Zehlicke T, Just T, Specht O, Schmidt W, Sass W, Moss C, Pau H-W

Аннотация

OBJECTIVE: Middle ear pressure depends widely on the function of the eustachian tube. Tube dysfunction is often a trigger for middle ear diseases like chronic otitis media but also for barotrauma. Patients with impaired tubal function should not be exposed to situations with extreme pressure changes. Until today, there is no valid examining method for long-time measuring of the development of middle ear pressure.

MATERIAL AND METHODS: The basic idea was to develop a thin flexible film with integrated strain gauges made of platinum and gold, which can be applied directly on the surface of the tympanic membrane. Using these, shifts or distensions of the tympanic membrane in a temporal bone model can be measured and documented.

RESULTS: We were able to show that the measuring system was sensitive enough to register pressure variations in the middle ear volume of a temporal bone model. While using a full bridge design of the strain gauge resistors it could be shown that the measuring voltage of the strain gauges were in phase of the pressure measurement curve or according to the positioning in opposite phase. In single resistor mode the measurement were not so positioning depended. Measuring tympanic membrane movement in case of perforation was feasible.

CONCLUSIONS: To improve the longtime stability of the strain gauge film the next development step will be to make a new design with Konstantan resistors (CU-Ni-Mn). After a testing period, longtime measurements in clinical studies will follow.