MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2016.08.030

Медицинская библиотека · Journal of cataract and refractive surgery · 2016

Toxic anterior segment syndrome caused by autoclave reservoir wall biofilms and their residual toxins.

Sorenson Andrew L, Sorenson Robert L, Evans David J

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To identify etiology of toxic anterior segment syndrome (TASS) after uneventful phacoemulsification.

SETTING

EyeMD Laser and Surgery Center, Oakland, California.

DESIGN

Retrospective case series.

METHODS

Patient charts with TASS were reviewed. Reservoirs of 2 autoclaves associated with these cases were cultured for bacterial contamination. Cultures were performed on 23 other autoclave reservoirs at surgery centers in the local area. The main outcome measures were the incidence of TASS and prevalence of bacterial biofilm contamination of autoclave reservoirs.

RESULTS

From 2010 to 2013, 11 935 consecutive cataract surgeries were performed at 1 center by multiple surgeons with no reported TASS. Between January 1, 2014, and January 15, 2015, 10 cases of TASS occurred out of 3003 cataract surgeries; these patients' charts were reviewed. Cultures of 2 Statim autoclave reservoir walls grew Bacillus species, Williamsia species, Mycobacterium mucogenicum, and Candida parapsilosis. Scanning electron microscopy of reservoir wall sections showed prominent biofilm. The 2 autoclaves were replaced in January 2015. Subsequently, 2875 cataract surgeries were performed with no reported TASS (P < .001, χ2 test). Eighteen of 23 additional regional autoclaves were also contaminated with bacterial biofilms.

CONCLUSIONS

Toxic anterior segment syndrome was strongly associated with bacterial biofilm contamination of autoclave reservoirs. An etiological mechanism might involve transport of heat-stable bacterial cell antigens in the steam with deposition on surgical instrumentation. Data suggest widespread prevalence of bacterial biofilms on fluid-reservoir walls, despite adherence to manufacturer guidelines for cleaning and maintenance. Prevention or elimination of autoclave fluid-reservoir biofilms might reduce the risk for postoperative TASS.

FINANCIAL DISCLOSURE: None of the authors has a financial or proprietary interest in any material or method mentioned.

Для цитирования: Sorenson Andrew L, Sorenson Robert L, Evans David J. Toxic anterior segment syndrome caused by autoclave reservoir wall biofilms and their residual toxins.. Journal of cataract and refractive surgery. 2016. DOI: 10.1016/j.jcrs.2016.08.030.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.