DOI 10.1016/j.clim.2017.05.007
An evaluation of the TREC assay with regard to the integration of SCID screening into the Dutch newborn screening program.
Аннотация
Newborn screening of severe combined immunodeficiency through the detection of T-cell receptor excision circles will provide the opportunity of treating before the occurrence of life-threatening infections. With the EnLite Neonatal TREC assay (PerkinElmer) and end-point PCR, 39 samples (3.0%) of 1295 heel prick cards of the Dutch newborn screening program required a retest after initial analysis. After retest, 21 samples (1.62%) gave TREC levels below cut-off. A significant reduction in TREC levels was observed in heel prick cards stored for three months (n=33) and one year (n=33). Preterm newborns (n=155) showed significantly lower TREC levels and a higher retest-rate than full-term newborns. Peripheral blood spots of 22 confirmed SCID patients and 17 primary immunodeficiency patients showed undetectable or low TREC-levels. These findings suggest that the EnLite Neonatal TREC assay is a suitable method for SCID-screening in the Netherlands, thereby providing guidance in the decisions concerning implementation into the Dutch program.