Comparison of artefact reduction possibilities with the new active transcutaneous bone conduction implant (Bonebridge).
Аннотация
OBJECTIVE
This study aimed to evaluate the possibilities of artefact reduction using different anatomical implant positions with the Bonebridge bone-conduction hearing implant 602 for a patient with an acoustic neuroma requiring regular diagnostic magnetic resonance imaging of the tumour position.
METHOD: Three implant positions and magnetic resonance imaging examinations with and without customised sequences for metal artefact suppression were investigated. The diagnostic usefulness was rated by a radiologist (qualitative evaluation), and the relation between the area of artefact and the total head area was calculated (quantitative evaluation).
RESULTS
Following the qualitative analysis, the radiologist rated the superior to middle fossa implant placement significantly better for diagnostic purposes, which is in agreement with the calculated artefact ratio (p < 0.0001). The customised slice-encoding metal artifact correction view-angle tilting metal artifact reduction technique sequences significantly decreased the relative artefact area between 5.13 per cent and 25.02 per cent. The smallest mean artefact diameter was found for the superior to middle fossa position with 6.80 ± 1.30 cm (range: 5.42-9.74 cm; reduction of 18.65 per cent).
CONCLUSION
The application of artefact reduction sequencing and special anatomical implant positioning allows regular magnetic resonance imaging in patients with the bone-conduction hearing implant 602 without sacrificing diagnostic imaging quality for tumour diagnosis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.