Intraocular Pressure Measurements After Descemet Membrane Endothelial Keratoplasty.

Maier Anna-Karina, Gundlach Enken, Pahlitzsch Milena, Gonnermann Johannes, Corkhill Caitlin, Bertelmann Eckart, Joussen Antonia M, Klamann Matthias K J, Torun Necip

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Intraocular pressure (IOP) elevation occurs regularly after Descemet membrane endothelial keratoplasty (DMEK). This study evaluated the effect of central corneal thickness (CCT) on the IOP after DMEK.

PATIENTS AND

METHODS

This prospective study recorded the IOP of 46 eyes from 46 patients preoperatively, and then 1 and 3 months after DMEK. IOP measurement was performed by noncontact pneumatic tonometry (NCT), iCare, Goldmann applanation tonometry (GAT), and dynamic contour tonometry (DCT). CCT was analyzed by anterior-segment optical coherence tomography.

RESULTS

Mean IOPs as measured by NCT, iCare, GAT, and DCT, respectively, were 14.3, 11.6, 12.7, and 16.0 mm Hg preoperatively; 13.1, 12.6, 12.7, and 16.0 mm Hg after 1 month; and 14.7, 14.5, 12.9, and 17.7 mm Hg after 3 months. There was a correlation of IOP measurements between GAT and NCT (P=0.119), GAT and iCare (P=0.892), and iCare and NCT (P=0.081) after 1 month and between iCare and NCT (P=0.702) after 3 months. Although GAT recorded approximately stable IOP values, NCT, iCare, and DCT tended to measure a higher IOP postoperatively than preoperatively. Preoperatively, correlations between IOP and CCT were not statistically significant for any measurement technique. After 3 months, the IOP measured by NCT and iCare correlated significantly with CCT (P=0.003, 0.041).

CONCLUSIONS

Correlation between the different measurement techniques was poor and showed a broad distribution of limits of agreement. Therefore, a change between the 4 techniques during follow-up is not recommended. Nevertheless, all techniques detected the IOP elevation during follow-up, requiring therapy.

Для цитирования: Maier Anna-Karina, Gundlach Enken, Pahlitzsch Milena, Gonnermann Johannes, Corkhill Caitlin, Bertelmann Eckart, Joussen Antonia M, Klamann Matthias K J, Torun Necip. Intraocular Pressure Measurements After Descemet Membrane Endothelial Keratoplasty.. Journal of glaucoma. 2017. DOI: 10.1097/ijg.0000000000000593.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.