Simulation of thoracic endovascular aortic repair in a perfused patient-specific model of type B aortic dissection.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mohl Lukas, Karl Roger, Hagedorn Matthias N, Runz Armin, Skornitzke Stephan, Toelle Malte, Bergt C Soeren, Hatzl Johannes, Uhl Christian, Böckler Dittmar, Meisenbacher Katrin, Engelhardt Sandy
Аннотация
PURPOSE: Complicated type B Aortic dissection is a severe aortic pathology that requires treatment through thoracic endovascular aortic repair (TEVAR). During TEVAR a stentgraft is deployed in the aortic lumen in order to restore blood flow. Due to the complicated pathology including an entry, a resulting dissection wall with potentially several re-entries, replicating this structure artificially has proven to be challenging thus far.
METHODS: We developed a 3d printed, patient-specific and perfused aortic dissection phantom with a flexible dissection flap and all major branching vessels. The model was segmented from CTA images and fabricated out of a flexible material to mimic aortic wall tissue. It was placed in a pulsatile hemodynamic flow loop. Hemodynamics were investigated through pressure and flow measurements and doppler ultrasound imaging. Surgeons performed a TEVAR intervention including stentgraft deployment under fluoroscopic guidance.
RESULTS: The flexible aortic dissection phantom was successfully incorporated in the hemodynamic flow loop, a systolic pressure of 112 mmHg and physiological flow of 4.05 L per minute was reached. Flow velocities were higher in true lumen with a up to 35.7 cm/s compared to the false lumen with a maximum of 13.3 cm/s, chaotic flow patterns were observed on main entry and reentry sights. A TEVAR procedure was successfully performed under fluoroscopy. The position of the stentgraft was confirmed using CTA imaging.
CONCLUSIONS: This perfused in-vitro phantom allows for detailed investigation of the complex inner hemodynamics of aortic dissections on a patient-specific level and enables the simulation of TEVAR procedures in a real endovascular operating environment. Therefore, it could provide a dynamic platform for future surgical training and research.
METHODS: We developed a 3d printed, patient-specific and perfused aortic dissection phantom with a flexible dissection flap and all major branching vessels. The model was segmented from CTA images and fabricated out of a flexible material to mimic aortic wall tissue. It was placed in a pulsatile hemodynamic flow loop. Hemodynamics were investigated through pressure and flow measurements and doppler ultrasound imaging. Surgeons performed a TEVAR intervention including stentgraft deployment under fluoroscopic guidance.
RESULTS: The flexible aortic dissection phantom was successfully incorporated in the hemodynamic flow loop, a systolic pressure of 112 mmHg and physiological flow of 4.05 L per minute was reached. Flow velocities were higher in true lumen with a up to 35.7 cm/s compared to the false lumen with a maximum of 13.3 cm/s, chaotic flow patterns were observed on main entry and reentry sights. A TEVAR procedure was successfully performed under fluoroscopy. The position of the stentgraft was confirmed using CTA imaging.
CONCLUSIONS: This perfused in-vitro phantom allows for detailed investigation of the complex inner hemodynamics of aortic dissections on a patient-specific level and enables the simulation of TEVAR procedures in a real endovascular operating environment. Therefore, it could provide a dynamic platform for future surgical training and research.