Pacing Using Cardiac Implantable Electric Device During TAVR: 10-Year Experience of a High-Volume Center.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Haum Magda, Steffen Julius, Sadoni Sebastian, Theiss Hans, Stark Konstantin, Estner Heidi, Massberg Steffen, Deseive Simon, Lackermair Korbinian
Аннотация
BACKGROUND: Transcatheter aortic valve replacement (TAVR) is an effective and safe therapy for severe aortic stenosis. Rapid or fast pacing is required for implantation, which can be performed via a pre-existing cardiac implantable electric device (CIED). However, safety data on CIEDs for pacing in TAVR are missing.
OBJECTIVES: The aim of this study was to elucidate procedural safety and feasibility of internal pacing with a CIED in TAVR.
METHODS: Patients undergoing TAVR with a CIED were included in this analysis. Baseline characteristics, procedural details, and complications according to Valve Academic Research Consortium 3 (VARC-3) criteria after TAVR were compared between both groups.
RESULTS: A total of 486 patients were included. Pacing was performed using a CIED in 150 patients and a transient pacemaker in 336 patients. No differences in technical success according to VARC-3 criteria or procedure duration occurred between the groups. The usage of transient pacers for pacing was associated with a significantly higher bleeding rate (bleeding type ≥2 according to VARC-3-criteria; 2.0% vs 13.1%; P < 0.01). Furthermore, impairment of the CIED appeared in 2.3% of patients after TAVR only in the group in which pacing was performed by a transient pacer, leading to surgical revision of the CIED in 1.3% of all patients when transient pacemakers were used.
CONCLUSIONS: Internal pacing using a CIED is safe and feasible without differences of procedural time and technical success and might reduce bleeding rates. Furthermore, pacing using a CIED circumvents the risk of lead dislocation. Our data provide an urgent call for the use of a CIED for pacing during a TAVR procedure in general.
OBJECTIVES: The aim of this study was to elucidate procedural safety and feasibility of internal pacing with a CIED in TAVR.
METHODS: Patients undergoing TAVR with a CIED were included in this analysis. Baseline characteristics, procedural details, and complications according to Valve Academic Research Consortium 3 (VARC-3) criteria after TAVR were compared between both groups.
RESULTS: A total of 486 patients were included. Pacing was performed using a CIED in 150 patients and a transient pacemaker in 336 patients. No differences in technical success according to VARC-3 criteria or procedure duration occurred between the groups. The usage of transient pacers for pacing was associated with a significantly higher bleeding rate (bleeding type ≥2 according to VARC-3-criteria; 2.0% vs 13.1%; P < 0.01). Furthermore, impairment of the CIED appeared in 2.3% of patients after TAVR only in the group in which pacing was performed by a transient pacer, leading to surgical revision of the CIED in 1.3% of all patients when transient pacemakers were used.
CONCLUSIONS: Internal pacing using a CIED is safe and feasible without differences of procedural time and technical success and might reduce bleeding rates. Furthermore, pacing using a CIED circumvents the risk of lead dislocation. Our data provide an urgent call for the use of a CIED for pacing during a TAVR procedure in general.