MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2018.03.023

Медицинская библиотека · JACC. Cardiovascular interventions · 2018

The SAVI-TF Registry: 1-Year Outcomes of the European Post-Market Registry Using the ACURATE neo Transcatheter Heart Valve Under Real-World Conditions in 1,000 Patients.

Kim Won-Keun, Hengstenberg Christian, Hilker Michael, Kerber Sebastian, Schäfer Ulrich, Rudolph Tanja, Linke Axel, Franz Norbert, Kuntze Thomas, Nef Holger, Kappert Utz, Zembala Michael O, Toggweiler Stefan, Walther Thomas, Möllmann Helge

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The SAVI-TF (Symetis ACURATE neo Valve Implantation Using Transfemoral Access) registry was initiated to study the ACURATE neo transcatheter heart valve in a large patient population treated under real-world conditions.

BACKGROUND

The self-expanding, supra-annular ACURATE neo prosthesis is a transcatheter heart valve that gained the Conformité Européene mark in 2014, but only limited clinical data are available so far.

METHODS

This prospective, multicenter registry enrolled 1,000 patients at 25 European centers who were followed for 1 year post-procedure.

RESULTS

Mean patient age was 81.1 ± 5.2 years; mean logistic European System for Cardiac Operative Risk Evaluation I score, European System for Cardiac Operative Risk Evaluation II score, and Society of Thoracic Surgeons score were 18.1 ± 12.5%, 6.6 ± 7.5%, and 6.0 ± 5.6%, respectively. At 1 year, 8.0% (95% confidence interval [CI]: 6.3% to 9.7%) of patients had died, 2.3% (95% CI: 1.3% to 3.2%) had disabling strokes, and 9.9% (95% CI: 8.1% to 11.8%) had permanent pacemaker implantations. Through 1 year, 5 reinterventions (0.5%; 95% CI: 0.1% to 1.0%) were performed: 3 valve-in-valve and 2 surgical aortic valve replacements. Mean effective orifice area was 1.84 ± 0.43 cm2, mean gradient was 7.3 ± 3.7 mm Hg, and greater than mild paravalvular leakage was observed in 3.6% of patients.

CONCLUSIONS

Transfemoral implantation of the ACURATE neo prosthesis resulted in favorable 1-year clinical and echocardiographic outcomes with very low mortality and new pacemaker rates.

Для цитирования: Kim Won-Keun, Hengstenberg Christian, Hilker Michael, Kerber Sebastian, Schäfer Ulrich, Rudolph Tanja, Linke Axel, Franz Norbert, Kuntze Thomas, Nef Holger, Kappert Utz, Zembala Michael O, Toggweiler Stefan, Walther Thomas, Möllmann Helge. The SAVI-TF Registry: 1-Year Outcomes of the European Post-Market Registry Using the ACURATE neo Transcatheter Heart Valve Under Real-World Conditions in 1,000 Patients.. JACC. Cardiovascular interventions. 2018. DOI: 10.1016/j.jcin.2018.03.023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.