Multicenter Evaluation of Prosthesis Oversizing of the SAPIEN 3 Transcatheter Heart Valve. Impact on Device Failure and New Pacemaker Implantations.

Pellegrini Costanza, Kim Won-Keun, Holzamer Andreas, Walther Thomas, Mayr N Patrick, Michel Jonathan, Rheude Tobias, Nuñez Julio, Kasel Albert M, Trenkwalder Teresa, Kaess Bernhard M, Joner Michael, Kastrati Adnan, Schunkert Heribert, Hilker Michael, Möllmann Helge, Hengstenberg Christian, Husser Oliver

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

INTRODUCTION AND

OBJECTIVES

A certain degree of prosthesis oversizing (OS) is recommended for the SAPIEN 3 to achieve device success. However, an increase in OS may increase permanent pacemaker implantation (PPI) rates. We therefore investigated the influence of OS on device failure and PPI.

METHODS

A total of 804 patients were treated with SAPIEN 3 at 3 centers. Multislice computed tomography-derived OS was calculated and analyzed both as a continuous variable and categorized in 5% increments with -4% to 0% as reference.

RESULTS

Device failure occurred in 8.8% of patients. Median OS was lower in patients with device failure vs those with device success (+4% vs +8%; P=.038). A nonlinear risk pattern was shown for OS with a significantly reduced device failure rate within 4% to +22% of OS. There was no case of paravalvular leakage II+ between +10% to +20% of OS. The overall PPI rate was 16.2% and the median OS was significantly larger in patients with PPI (PPI: +9% vs no PPI: +7%; P = .025), while implantation depth did not vary in patients with vs without PPI (6.9±1.7 mm vs 6.6±1.9 mm; P=.101). The risk of PPI increased with increasing OS and was highest in the 2 highest categories.

CONCLUSIONS

An increase in OS reduces the risk for device failure but increases the risk for PPI. There was no ideal range of OS to minimize both device failure and PPI.

Для цитирования: Pellegrini Costanza, Kim Won-Keun, Holzamer Andreas, Walther Thomas, Mayr N Patrick, Michel Jonathan, Rheude Tobias, Nuñez Julio, Kasel Albert M, Trenkwalder Teresa, Kaess Bernhard M, Joner Michael, Kastrati Adnan, Schunkert Heribert, Hilker Michael, Möllmann Helge, Hengstenberg Christian, Husser Oliver. Multicenter Evaluation of Prosthesis Oversizing of the SAPIEN 3 Transcatheter Heart Valve. Impact on Device Failure and New Pacemaker Implantations.. Revista espanola de cardiologia (English ed.). 2018. DOI: 10.1016/j.rec.2018.06.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.