MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10549-018-4878-8

Медицинская библиотека · Breast cancer research and treatment · 2018

Dosimetric study to assess the feasibility of intraoperative radiotherapy with electrons (ELIOT) as partial breast irradiation for patients with cardiac implantable electronic device (CIED).

Luraschi Rosa, Lazzari Roberta, Galimberti Viviana, Rondi Elena, Bazani Alessia, Corso Giovanni, Colombo Nicola, Ricotti Rosalinda, Fodor Cristiana, Winnicki Mikolaj, Leonardi Maria Cristina, Jereczek-Fossa Barbara Alicja, Cattani Federica

Клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To report in-vivo dosimetry in the infraclavicular region, a potential site of a cardiac implantable electronic device (CIED) and to evaluate the absorbed dose from intraoperative radiotherapy with electrons (ELIOT).

METHODS

27 non-cardiopathic breast cancer (BC) patients without CIED received quadrantectomy and ELIOT as partial breast irradiation. Before delivering ELIOT, two catheters, each containing eight thermoluminescent dosimeters (TLDs), were positioned in the infraclavicular region. TLDs internal catheter was located deep in the tumor bed while the external catheter was placed on patient's skin.

RESULTS

Data were available for 24/27 patients. The absorbed doses were referred to the dose of 21 Gy. Values measured by the external catheter were low, although statistically significant higher doses were found close to the applicator (mean values 0.26-0.49 Gy). External TLD doses in proximity of the applicator were lower than those detected by their internal counterparts. Values measured by the internal catheter TLDs varied according to the distance from the applicator while no correlation with tumor site and beam energy was found. The distance from the applicator to deliver < 2 Gy to a CIED was 2 cm, while from 2.5 cm the dose measured in all the patients became negligible.

CONCLUSIONS

This dosimetric study provided data to support the clinical use of ELIOT in BC patients having CIEDs as long as the suggested minimum safe distance of 2.5 cm is taken from the RT field in case of ELIOT single dose of 21 Gy, in the energy range of 6-10 MeV.

Для цитирования: Luraschi Rosa, Lazzari Roberta, Galimberti Viviana, Rondi Elena, Bazani Alessia, Corso Giovanni, Colombo Nicola, Ricotti Rosalinda, Fodor Cristiana, Winnicki Mikolaj, Leonardi Maria Cristina, Jereczek-Fossa Barbara Alicja, Cattani Federica. Dosimetric study to assess the feasibility of intraoperative radiotherapy with electrons (ELIOT) as partial breast irradiation for patients with cardiac implantable electronic device (CIED).. Breast cancer research and treatment. 2018. DOI: 10.1007/s10549-018-4878-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.