DOI 10.1016/j.jvir.2019.08.036

Медицинская библиотека · Journal of vascular and interventional radiology : JVIR · 2019

Final Two-Year Outcomes for the Sentry Bioconvertible Inferior Vena Cava Filter in Patients Requiring Temporary Protection from Pulmonary Embolism.

Dake Michael D, Murphy Timothy P, Krämer Albrecht H, Darcy Michael D, Sewall Luke E, Curi Michael A, Johnson Matthew S, Arena Frank, Swischuk James L, Ansel Gary M, Silver Mitchell J, Saddekni Souheil, Brower Jayson S, Mendes Robert, SENTRY Trial Investigators

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To report final 2-year outcomes with the Sentry bioconvertible inferior vena cava (IVC) filter in patients requiring temporary protection against pulmonary embolism (PE).

MATERIALS AND

METHODS

In a prospective multicenter trial, the Sentry filter was implanted in 129 patients with documented deep vein thrombosis (DVT) and/or PE (67.5%) or who were at temporary risk of developing DVT/PE (32.6%). Patients were monitored and bioconversion status ascertained by radiography, computed tomography (CT), and CT venography through 2 years.

RESULTS

The composite primary 6-month endpoint of clinical success was achieved in 97.4% (111/114) of patients. The rate of new symptomatic PE was 0% (n = 126) through 1 year and 2.4% (n = 85) through the second year of follow-up, with 2 new nonfatal cases at 581 and 624 days that were adjudicated as not related to the procedure or device. Two patients (1.6%) developed symptomatic caval thrombosis during the first month and underwent successful interventions without recurrence. No other filter-related symptomatic complications occurred through 2 years. There was no filter tilting, migration, embolization, fracture, or caval perforation and no filter-related deaths through 2 years. Filter bioconversion was successful for 95.7% (110/115) of patients at 6 months, 96.4% (106/110) of patients at 12 months, and 96.5% (82/85) of patients at 24 months. Through 24 months of follow-up, there was no evidence of late-stage IVC obstruction or thrombosis after filter bioconversion or of thrombogenicity associated with retracted filter arms.

CONCLUSIONS

The Sentry IVC filter provided safe and effective protection against PE, with a high rate of intended bioconversion and a low rate of device-related complications, through 2 years of follow-up.

Для цитирования: Dake Michael D, Murphy Timothy P, Krämer Albrecht H, Darcy Michael D, Sewall Luke E, Curi Michael A, Johnson Matthew S, Arena Frank, Swischuk James L, Ansel Gary M, Silver Mitchell J, Saddekni Souheil, Brower Jayson S, Mendes Robert, SENTRY Trial Investigators. Final Two-Year Outcomes for the Sentry Bioconvertible Inferior Vena Cava Filter in Patients Requiring Temporary Protection from Pulmonary Embolism.. Journal of vascular and interventional radiology : JVIR. 2019. DOI: 10.1016/j.jvir.2019.08.036.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.