Pourbiax Diagrams as an Aid to Understanding the Impact of Acid/Base Disturbance on Blood Glucose Point-of-Care Testing.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
McPherson Peter A C, McClements Owen S, Johnston Ben M
Аннотация
OBJECTIVE: Assays based on redox reactions that involve proton transfer are vulnerable to artifactual findings in metabolic acidosis/alkalosis. We evaluated the impact of pH on the measurement of blood glucose by the glucose dehydrogenase/pyrroloquinoline quinone system used in point-of-care-testing.
METHODS: We applied a series of thermodynamic equations to adjust the Gibbs energy for the pyrroloquinoline quinone couple. This adjusts values taken under standard conditions to those more closely resembling the physiological state.
RESULTS: Under standard conditions, the pyrroloquinoline quinone couple has Eo = -0.125 V whereas adjustment to the physiological state (pH 7.40, ionic strength 0.15 mol/L, and temperature 310.15°K) yields Eo' = -0.166 V. This corresponds to an uncertainty in blood glucose determination of approximately 0.13 mmol/L.
CONCLUSION: We have demonstrated that the impact of pH on blood glucose determination by the glucose dehydrogenase/pyrroloquinoline quinone system (under physiologically relevant conditions of ionic strength and temperature) is not clinically significant.
METHODS: We applied a series of thermodynamic equations to adjust the Gibbs energy for the pyrroloquinoline quinone couple. This adjusts values taken under standard conditions to those more closely resembling the physiological state.
RESULTS: Under standard conditions, the pyrroloquinoline quinone couple has Eo = -0.125 V whereas adjustment to the physiological state (pH 7.40, ionic strength 0.15 mol/L, and temperature 310.15°K) yields Eo' = -0.166 V. This corresponds to an uncertainty in blood glucose determination of approximately 0.13 mmol/L.
CONCLUSION: We have demonstrated that the impact of pH on blood glucose determination by the glucose dehydrogenase/pyrroloquinoline quinone system (under physiologically relevant conditions of ionic strength and temperature) is not clinically significant.