Performance of home-based self-tonometry (iCare HOME (TA022)) for measuring intraocular pressure among healthy and glaucoma patients.

Nayak Anush, Ramesh S Ve, Kuzhuppilly Neetha I R, Pai Vijaya H, Chaitanya Aditya

СтатьяПеревод пока не подготовлен

Аннотация

Introduction

The purpose of this study was to compare iCare HOME (TA022) with Goldmann applanation tonometer and to evaluate the self-tonometry measurements among the Indian population.

Methods

Eligible patients underwent iCare HOME training through guided demonstration (verbal, pictorial, video) and practised self-tonometry measures using iCare HOME. Certification for independent iCare HOME measure was provided if first iCare HOME intraocular pressure (IOP) measurement fell within ± 5 mmHg of Goldmann applanation tonometer (GAT) measurement which was measured by the trained clinician (principal investigator). Certified participants underwent simulated home self-tonometry measurements using iCare HOME, and agreement with GAT IOP measurements was assessed.

Results

Seven of 83 participants (8.43%) failed to complete the study due to difficulty in performing the task, leading to non-certification. Patients who could use the iCare HOME had a mean age of 53 ± 15.55years (53% males; 46% females). Only one in 12 subjects did not qualify to use iCare HOME. The overall mean difference between iCare HOME and GAT was 0.83 mmHg (95%, 3.92 and -2.25). At various pressure ranges, 7-16 mmHg, 17-23 mmHg and >23 mmHg, the mean difference between iCare HOME and GAT was 1.22 mmHg (95%, 4.32 and -1.86), 0.77 mmHg (95%,3.69 and -2.19), -0.11 mmHg (95%, 2.52 and -2.74) respectively. The intra-class correlation coefficient of the iCare HOME device was 0.997(95% CI,0.995-0.998).

Conclusions

Patients were able to perform self- tonometry using iCare HOME with good reliability and safety. iCare HOME can be used to address the issue of difficulty in acquiring frequent and diurnal IOP measurements by patients doing self-tonometry from home.

Для цитирования: Nayak Anush, Ramesh S Ve, Kuzhuppilly Neetha I R, Pai Vijaya H, Chaitanya Aditya. Performance of home-based self-tonometry (iCare HOME (TA022)) for measuring intraocular pressure among healthy and glaucoma patients.. F1000Research. 2023. DOI: 10.12688/f1000research.123104.2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.