Gastric cancer : official journal of the International Gastric Cancer Association and the Japanese Gastric Cancer Association · 2022
Feasibility and safety of inserting transient biodegradable stents in the pylorus during pylorus-preserving gastrectomy for gastric cancer: a preliminary study in a porcine for proof of concept.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Park Ji-Hyeon, Yoon Hyesung, Kwak Yoon Jin, Wang Chaojie, Alzahrani Khalid Mohammed, Wang Sen, Alzahrani Fadhel Dhaifallah H, Kim Hyun Myong, Koo Eunhee, Yoo Ja Eun, Choi Jong-Ho, Park Shin-Hoo, Kong Seong-Ho, Park Do Joong, Lee Hyuk-Joon, Yang Han-Kwang
Аннотация
BACKGROUND: To evaluate whether insertion of self-biodegradable stent into the pylorus to prevent delayed-gastric emptying after pylorus-preserving gastrectomy is feasible and safe through porcine experiment.
METHODS: Self-biodegradable dumbbell-shaped pyloric stents were designed from absorbable suture materials: poly(glycolide-co-caprolactone) (PGCL) or poly-p-dioxanone (PPDO). After gastrotomy on ten pigs, each stent was inserted: two shams, four PGCL stents, and four PPDO stents. Body weight (Bwt), body temperature (BT), complete blood cell (CBC) count, and plain X-ray were evaluated. On postoperative day (POD) 13, euthanasia was performed for histologic evaluation.
RESULTS: Operation was successfully performed in all ten pigs. Without tagging suture, both stents migrated before POD 3. The migration was delayed up to POD 13, when the tagging sutures (-t) were applied between stent and stomach wall. Self-degradation of PGCL started from POD 3, and stents were completely excreted from the abdomen by POD 8. Although PPDO were also weakened as self-degradation progressed, its shape was maintained in gastrointestinal tract for 13 days. Unexpected sudden death occurred in the pig with PPDO-t2 on POD 10, which is more likely due to acute volvulus rather than stent-related complication. There was no significant difference between three groups in terms of Bwt, BT, CBC, and histology (sham vs. PGCL vs. PPDO, all p > 0.05).
CONCLUSION: The concept of biodegradable stents made of absorbent suture material seems feasible in porcine experiment. Among them, PGCL which has shown rapid absorption, appears to be a more suitable material for transient pyloric absorbable stent when considering safety aspect.
METHODS: Self-biodegradable dumbbell-shaped pyloric stents were designed from absorbable suture materials: poly(glycolide-co-caprolactone) (PGCL) or poly-p-dioxanone (PPDO). After gastrotomy on ten pigs, each stent was inserted: two shams, four PGCL stents, and four PPDO stents. Body weight (Bwt), body temperature (BT), complete blood cell (CBC) count, and plain X-ray were evaluated. On postoperative day (POD) 13, euthanasia was performed for histologic evaluation.
RESULTS: Operation was successfully performed in all ten pigs. Without tagging suture, both stents migrated before POD 3. The migration was delayed up to POD 13, when the tagging sutures (-t) were applied between stent and stomach wall. Self-degradation of PGCL started from POD 3, and stents were completely excreted from the abdomen by POD 8. Although PPDO were also weakened as self-degradation progressed, its shape was maintained in gastrointestinal tract for 13 days. Unexpected sudden death occurred in the pig with PPDO-t2 on POD 10, which is more likely due to acute volvulus rather than stent-related complication. There was no significant difference between three groups in terms of Bwt, BT, CBC, and histology (sham vs. PGCL vs. PPDO, all p > 0.05).
CONCLUSION: The concept of biodegradable stents made of absorbent suture material seems feasible in porcine experiment. Among them, PGCL which has shown rapid absorption, appears to be a more suitable material for transient pyloric absorbable stent when considering safety aspect.