Comparison of immulite with vidas for detection of infection in a low-prevalence population of pregnant women in The Netherlands.

Vlaspolder F, Singer P, Smit A, Diepersloot R J

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

A comparative evaluation of the Vidas system (bioMérieux, Marcy l'Etoile, France) and the Immulite System (Diagnostic Products Corporation) was performed using 500 prospectively collected serum samples. As part of a routine antenatal screening program, these samples were tested for hepatitis B surface antigen, and immunoglobulin G (IgG) and IgM antibodies to Toxoplasma gondii and rubella virus. The overall agreement between the two assay systems ranged from 98.0 to 99.8%. After discrepancy analysis the outcome in terms of relative sensitivity and specificity varied from 97.5 to 100%.

Для цитирования: Vlaspolder F, Singer P, Smit A, Diepersloot R J. Comparison of immulite with vidas for detection of infection in a low-prevalence population of pregnant women in The Netherlands.. Clinical and diagnostic laboratory immunology. 2001. DOI: 10.1128/cdli.8.3.552-555.2001.