Comparison of corneal pachymetry evolution after accelerated corneal crosslinking in keratoconus eyes using anterior segment optical coherence tomography and Scheimpflug imaging.

Lang Antonia, Seitz Berthold, Munteanu Cristian, Flockerzi Elias

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: This retrospective clinical study compared longitudinal corneal pachymetry measurement evolution after accelerated corneal crosslinking (A-CXL, 9 mW/cm2, 10 min) in keratoconus (KC) eyes as measured by Pentacam HR Scheimpflug imaging (Oculus, Germany) and Casia 2 (Tomey, Japan) anterior segment optical coherence tomography (AS-OCT).

METHODS

218 eyes covering the full topographic KC (TKC) severity spectrum that underwent A-CXL between 2017 and 2021 were examined at five time points (before, 6 weeks, 6 months, 1 year after A-CXL and at the latest follow-up) using both Pentacam HR and Casia 2. Central (CCT) and thinnest (TCT) corneal thickness were assessed. Corneal densitometry and TKC classification were assessed by Pentacam HR only. Intraoperative CCT evolution was monitored using ultrasound (Pachymeter SP-3000, Tomey, Japan) to ensure that CCT did not fall below 400 µm.

RESULTS

CCT and TCT measured by Pentacam HR were significantly lower than preoperatively during the first postoperative year, while Casia 2 CCT and TCT measurements were not significantly lower than preoperatively at any time. The highest agreement of Pentacam HR and Casia 2 measurements occurred six weeks after A-CXL. Corneal densitometry peaked six weeks after A-CXL and decreased thereafter without reaching pre-A-CXL values. Ultrasound revealed significant CCT thinning during A-CXL.

CONCLUSION

Corneal Scheimpflug and AS-OCT based thickness measurements are not interchangeable. Corneal thinning after A-CXL detected by Pentacam HR was not observed in Casia 2 measurements. Since higher densitometry values did not correlate with a greater inter-device measurement difference, Scheimpflug-based corneal thickness measurements should be interpreted cautiously during the first year after A-CXL.

Для цитирования: Lang Antonia, Seitz Berthold, Munteanu Cristian, Flockerzi Elias. Comparison of corneal pachymetry evolution after accelerated corneal crosslinking in keratoconus eyes using anterior segment optical coherence tomography and Scheimpflug imaging.. International ophthalmology. 2025. DOI: 10.1007/s10792-025-03763-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.