Diagnostic evaluation of an in-house developed single-tube, duplex, nested IS6110 real-time PCR assay for rapid pulmonary tuberculosis diagnosis.

Leung Kenneth Siu-Sing, Siu Gilman Kit-Hang, Tam Kingsley King-Gee, Ho Pak-Leung, Wong Samson Sai-Yin, Leung Eunice Ka-Chun, Yu Shi Hui, Ma Oliver Chiu-Kit, Yam Wing-Cheong

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To perform a prospective evaluation on the diagnostic performance of an in-house developed, duplex nested IS6110 real-time Polymerase-Chain-Reaction (PCR) assay (IS6110-qPCR assay) for rapid pulmonary TB diagnosis.

METHODS

A total of 503 sputum specimens were prospectively collected from July 2016 to November 2016. Diagnostic accuracy and optimal cut-off Cycle-threshold (Ct) value for IS6110-qPCR assay was determined by Receiver Operating Characteristic (ROC) curve. Using the optimal cut-off Ct, diagnostic performance of IS6110-qPCR assay was assessed with reference to both bacteriological and clinical information. Meanwhile, limit of detection (LOD) was calculated using Mycobacterium tuberculosis H37Rv as reference strain.

RESULT: ROC curve analysis of IS6110-qPCR assay showed a high Area Under the Curve (AUC) value (0.948) with optimal Ct value at 24.140. Prospective analysis of IS6110-qPCR assay with cut-off Ct = 24.140 showed a high overall sensitivity and specificity of 97.2% and 99.7%, respectively. No cross reactivity was observed among all non-tuberculous mycobacteria specimens in this study. LOD analysis on MTB-spiked sputum showed an average detection limit of 5.0 CFU/mL at Ct = 23.18 (±SD, 0.57).

CONCLUSION

IS6110-qPCR assay is a highly accurate and cost-effective assay developed for primary screening of suspected TB cases, which is particularly suitable for regions with limited resources but high TB burden.

Для цитирования: Leung Kenneth Siu-Sing, Siu Gilman Kit-Hang, Tam Kingsley King-Gee, Ho Pak-Leung, Wong Samson Sai-Yin, Leung Eunice Ka-Chun, Yu Shi Hui, Ma Oliver Chiu-Kit, Yam Wing-Cheong. Diagnostic evaluation of an in-house developed single-tube, duplex, nested IS6110 real-time PCR assay for rapid pulmonary tuberculosis diagnosis.. Tuberculosis (Edinburgh, Scotland). 2018. DOI: 10.1016/j.tube.2018.08.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.