DOI 10.3928/1081-597x-20020901-27
Adaptive optics visual simulator.
Аннотация
PURPOSE: To develop a prototype instrument that uses adaptive optics to introduce virtually any desired aberration profile in a subject's eye. At the same time, the instrument could be used to evaluate the subject's spatial vision for each controlled aberration profile. This "aberration testing station" or "visual simulator" allows us to study the relationship between specific aberrations and visual quality.
METHODS
The apparatus uses infrared light to measure the wavefront aberration of the system plus the eye with a Hartmann-Shack wavefront sensor. Defocus is added (or removed) with a computer-controlled, motorized optometer, while higher order aberrations are introduced by a 37-channel membrane deformable mirror. A parallel viewing channel is used for visual testing with the instrument. Visual acuity, contrast sensitivity, and other visual tests are performed under normal viewing for each desired aberration profile.
RESULTS
The range of defocus that can be added is nearly unlimited, while the maximum amount of other aberration modes is restricted to approximately 0.5 microm, depending on mode. Pure modes or any selected combination of modes can be produced with high repeatability and precision (usually better than 0.05 microm), and the system works for pupil diameters up to 6 mm (with a natural pupil).
CONCLUSIONS
The adaptive optics visual simulator is a powerful, non-invasive tool to evaluate how aberrations affect vision. In addition, it can be used for the interactive design and testing of new ophthalmic devices, and for the simulation of visual outcomes in customized refractive surgery.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.