Fluid-structure interaction study on the causes of mending material damage after sigmoid sinus wall reconstruction.

Xu Kaihang, Qiu Xiaoyu, Dai Chihang, He Kaixuan, Wang Guopeng, Mu Zhenxia, Gao Bin, Gong Shusheng, Wang Zhenchang, Zhao Pengfei

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVE

Sigmoid Sinus (SS) Wall Reconstruction (SSWR) is the mainstream treatment for pulsatile tinnitus (PT), but it has a high risk of recurrence. The damage of mending material is the key cause of recurrence, and its hemodynamic mechanism is still unclear. The purpose of this study was to investigate the hemodynamic causes of mending material breakage.

METHODS

In this study, six patient-specific geometric models were reconstructed based on the data of the computed tomography angiography (CTA). The transient fluid-structure coupling method was performed to clarify the hemodynamic state of sigmoid sinus and the biomechanical state of the mending material. The distribution of stress and displacement and the flow pattern were calculated to evaluate the hemodynamic and biomechanics difference at the mending material area.

RESULTS

The area of blood flow impact in some patients (2/6) was consistent with the damaged location of the mending material. The average stress (6/6) and average displacement (6/6) of damaged mending material were higher than those of complete mending material. All (6/6) patients showed that the high-stress and high-displacement proportion of the DMM region was higher than that of the CMM region. Moreover, the average stress fluctuation (6/6) and average displacement (6/6) fluctuation degree of damaged mending material is larger than that of complete mending material.

CONCLUSIONS

The impact of blood and the uneven stress and displacement fluctuation of the mending material may be the causes of mending material damage. High stress and high displacement might be the key causes of the mending material damage.

Для цитирования: Xu Kaihang, Qiu Xiaoyu, Dai Chihang, He Kaixuan, Wang Guopeng, Mu Zhenxia, Gao Bin, Gong Shusheng, Wang Zhenchang, Zhao Pengfei. Fluid-structure interaction study on the causes of mending material damage after sigmoid sinus wall reconstruction.. Computer methods and programs in biomedicine. 2024. DOI: 10.1016/j.cmpb.2024.108040.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.