External outflow graft obstruction (OGO) in durable left ventricular assist devices (LVADs) is a rare but critical complication. Detecting external OGO can be challenging because of its nonspecific symptoms and a wide variety of differential diagnoses. We report the case of a 45 year old man with LVAD admitted with complaints of recurrent syncope and frequent low-flow alarms. Laboratory tests and chest radiography revealed no specific findings. Transthoracic echocardiography via the left subcostal approach revealed an outflow graft compressed by a low-echo area within a bend relief. This finding led to a diagnosis of external OGO, which was consistent with what was observed on contrast-enhanced computed tomography. The patient underwent surgery and the gelatinous substance compressing the outflow graft was successfully removed. This case highlights the utility of transthoracic echocardiography as a noninvasive and effective method for detecting external OGOs, emphasizing its role as a valuable tool for the initial screening of patients with suspected OGO.

Для цитирования: Nagai Shin, Komeyama Shotaro, Kondo Toru, Morimoto Ryota, Kazama Shingo, Hiraiwa Hiroaki, Okumura Takahiro, Yoshizumi Tomo, Mutsuga Masato, Murohara Toyoaki. Direct Visualization of External Outflow Graft Obstruction in Durable Left Ventricular Assist Device Using Transthoracic Echocardiography.. ASAIO journal (American Society for Artificial Internal Organs : 1992). 2025. DOI: 10.1097/mat.0000000000002406.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.