Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients.

Wörmann Jonas, Strik Marc, Jurisic Stjepan, Stout Kara, Elrefai Mohamed, Becher Nina, Schaer Beat, van Stipdonk Antonius, Srinivasan Neil T, Ploux Sylvain, Breitenstein Alexander, Kron Jordana, Roberts Paul R, Toennis Tobias, Linz Dominik, Dulai Rajdip, Hermes-Laufer Julia, Koneru Jay, Erküner Ömer, Dittrich Sebastian, van den Bruck Jan-Hendrik, Schipper Jan-Hendrik, Sultan Arian, Rosenberger Kerstin D, Steven Daniel, Lüker Jakob

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

AIMS: Subcutaneous implantable cardioverter-defibrillators (S-ICDs) have become established in preventing sudden cardiac death, with some advantages over transvenous defibrillator systems, including a lower incidence of lead failures. Despite technological advancements, S-ICD carriers may suffer from significant complications, such as premature battery depletion (PBD), that led to an advisory for nearly 40 000 patients. This multicentre study evaluated the incidence of PBD in a large set of S-ICD patients.

METHODS AND

RESULTS

Data from patients implanted with S-ICD models A209 and A219 between October 2012 and July 2023 across nine centres in Europe and the USA were reviewed. Incidence and implications of PBD, defined as clinically observed sudden drop in battery longevity, were analysed and compared to PBD with the definition of battery depletion within 60 months. Prospectively collected clinical data were obtained retrospectively from medical records, device telemetry, and manufacturer reports. This registry is listed on ClinicalTrials.gov (NCT05713708). Of the 1112 S-ICD devices analysed, 547 (49.2%) were equipped with a potentially affected capacitor linked to PBD occurrence, currently under Food and Drug Administration advisory. The median follow-up time for all patients was 46 [inter-quartile range (IQR) 24-63] months. Clinically suspected PBD was observed in 159 (29.1%) of cases, with a median time to generator removal or replacement of 65 (IQR 55-72) months, indicative of significant deviations from expected battery lifespan. Manufacturer confirmation of PBD was made in 91.7% of devices returned for analysis. No cases of PBD were observed in devices that were not under advisory.

CONCLUSION

This manufacturer-independent analysis highlights a notable incidence of PBD in patients equipped with S-ICD models under advisory, and the rate of PBD in this study corresponds to the rate currently estimated by the manufacturer. To the best of our knowledge, this provides the largest contemporary peer-reviewed study cohort investigating the actual incidence of PBD in S-ICD patients. These findings emphasize the importance of post-market registries in collaboration between clinicians and the manufacturer to optimize safety and efficacy in S-ICD treatment.

Для цитирования: Wörmann Jonas, Strik Marc, Jurisic Stjepan, Stout Kara, Elrefai Mohamed, Becher Nina, Schaer Beat, van Stipdonk Antonius, Srinivasan Neil T, Ploux Sylvain, Breitenstein Alexander, Kron Jordana, Roberts Paul R, Toennis Tobias, Linz Dominik, Dulai Rajdip, Hermes-Laufer Julia, Koneru Jay, Erküner Ömer, Dittrich Sebastian, van den Bruck Jan-Hendrik, Schipper Jan-Hendrik, Sultan Arian, Rosenberger Kerstin D, Steven Daniel, Lüker Jakob. Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2024. DOI: 10.1093/europace/euae170.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.