[The impact of implantable cardioverter defibrillator with lead alert function on inappropriate shocks caused by lead malfunctions].

Yuan C Z, Ze F, Wu C C, Duan J B, Zhou X, Yang D D, Li D, Li X B

Аннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

Objective

To evaluate the incidence and frequency of inappropriate shocks caused by defibrillation lead failure in patients with implantable cardioverter-defibrillators (ICD), and to explore methods for reducing the incidence and frequency of such inappropriate shocks.

Methods

This was a single-center retrospective study involving patients treated for defibrillation lead failures at Peking University People's Hospital between March 2015 and May 2024. Patients were divided into an alarm function group and a non-alarm function group based on whether their ICDs were equipped with lead alarm functions. Clinical data, lead data, and the incidence and frequency of inappropriate shocks were collected and compared between the two groups. A multivariate logistic regression model was used to analyze factors influencing the incidence and frequency of inappropriate ICD shocks. Kaplan-Meier survival curves were plotted to compare the trends in the incidence and frequency of inappropriate shocks over time since ICD implantation between the two groups.

Results

A total of 59 patients were enrolled, with a age of (56.7±15.2) years, including 42 males (71%). The lifespan of the failed leads in the entire cohort was 64.0 (36.0, 96.0) months. There were 26 patients in the alarm function group and 33 in the non-alarm function group. The most common manifestations of lead failure were oversensing (85%, 50/59) and abnormal pacing impedance (42%, 25/59). A total of 33 patients (56%, 33/59) experienced inappropriate shock therapy, with an average of 27.3 shocks per patient. The frequency of inappropriate ICD shocks in the non-alarm function group was higher than that in the alarm function group (25.0 (10.0, 60.0) times/year vs. 5.0 (2.8, 7.8) times/year, P=0.001). Multivariate logistic regression analysis showed that oversensing (OR=2.057, 95%CI 1.125-6.763, P=0.019) was an influencing factor for incidence of inappropriate shocks, while the lead alert function (OR=0.062, 95%CI 0.005-0.719, P=0.001) was a factor influencing the frequency of inappropriate shocks. Kaplan-Meier survival analysis revealed that the incidence and frequency of inappropriate shocks increased with the duration of ICD implantation in both groups, but the differences were not statistically significant (incidence: log-rank P=0.908; frequency: log-rank P=0.767).

Conclusion

The lead alert function can reduce the frequency of inappropriate shocks caused by lead failure.

Для цитирования: Yuan C Z, Ze F, Wu C C, Duan J B, Zhou X, Yang D D, Li D, Li X B. [The impact of implantable cardioverter defibrillator with lead alert function on inappropriate shocks caused by lead malfunctions].. Zhonghua xin xue guan bing za zhi. 2025. DOI: 10.3760/cma.j.cn112148-20251014-00720.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.