Complications of implantable cardioverter-defibrillator treatment in arrhythmogenic right ventricular cardiomyopathy.

Christensen Alex Hørby, Platonov Pyotr G, Svensson Anneli, Jensen Henrik K, Rootwelt-Norberg Christine, Dahlberg Pia, Madsen Trine, Frederiksen Tanja Charlotte, Heliö Tiina, Haugaa Kristina H, Bundgaard Henning, Svendsen Jesper H

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Treatment with implantable cardioverter-defibrillators (ICD) is a cornerstone for prevention of sudden cardiac death in arrhythmogenic right ventricular cardiomyopathy (ARVC). We aimed at describing the complications associated with ICD treatment in a multinational cohort with long-term follow-up.

METHODS AND

RESULTS

The Nordic ARVC registry was established in 2010 and encompasses a large multinational cohort of ARVC patients, including their clinical characteristics, treatment, and events during follow-up. We included 299 patients (66% males, median age 41 years). During a median follow-up of 10.6 years, 124 (41%) patients experienced appropriate ICD shock therapy, 28 (9%) experienced inappropriate shocks, 82 (27%) had a complication requiring surgery (mainly lead-related, n = 75), and 99 (33%) patients experienced the combined endpoint of either an inappropriate shock or a surgical complication. The crude rate of first inappropriate shock was 3.4% during the first year after implantation but decreased after the first year and plateaued over time. Contrary, the risk of a complication requiring surgery was 5.5% the first year and remained high throughout the study period. The combined risk of any complication was 7.9% the first year. In multivariate cox regression, presence of atrial fibrillation/flutter was a risk factor for inappropriate shock (P < 0.05), whereas sex, age at implant, and device type were not (all P > 0.05).

CONCLUSION

Forty-one percent of ARVC patients treated with ICD experienced potentially life-saving ICD therapy during long-term follow-up. A third of the patients experienced a complication during follow-up with lead-related complications constituting the vast majority.

Для цитирования: Christensen Alex Hørby, Platonov Pyotr G, Svensson Anneli, Jensen Henrik K, Rootwelt-Norberg Christine, Dahlberg Pia, Madsen Trine, Frederiksen Tanja Charlotte, Heliö Tiina, Haugaa Kristina H, Bundgaard Henning, Svendsen Jesper H. Complications of implantable cardioverter-defibrillator treatment in arrhythmogenic right ventricular cardiomyopathy.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2022. DOI: 10.1093/europace/euab112.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.