MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/0885066610392516

Медицинская библиотека · Journal of intensive care medicine · 2011

Massive aspiration past the tracheal tube cuff caused by closed tracheal suction system.

Dave Mital H, Frotzler Angela, Madjdpour Caveh, Koepfer Nelly, Weiss Markus

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Aspiration past the tracheal tube cuff has been recognized to be a risk factor for the development of ventilator-associated pneumonia (VAP). This study investigated the effect of closed tracheal suctioning on aspiration of fluid past the tracheal tube cuff in an in vitro benchtop model.

METHODS

High-volume low pressure tube cuffs of 7.5 mm internal diameter (ID) were placed in a 22 mm ID artificial trachea connected to a test lung. Positive pressure ventilation (PPV) with 15 cm H₂O peak inspiratory pressure and 5 cm H₂O positive end-expiratory pressure (PEEP) was used. A closed tracheal suction system (CTSS) catheter (size 14Fr) was attached to the tracheal tube and suction was performed for 5, 10, 15, or 20 seconds under 200 or 300 cm H₂O suction pressures. Amount of fluid (mL) aspirated along the tube cuff and the airway pressure changes were recorded for each suction procedure. Fluid aspiration during different suction conditions was compared using Kruskal-Wallis and Mann-Whitney test (Bonferroni correction [α = .01]).

RESULTS

During 10, 15, and 20 seconds suction, airway pressure consistently dropped down to -8 to -13 cm H₂O (P < .001) from the preset level. Fluid aspiration was never observed under PPV + PEEP but occurred always during suctioning. Aspiration along the tube cuff was higher with -300 cm H₂O than with -200 cm H₂O suction pressure (P < .001) and was much more during 15 and 20 seconds suction time as compared to 5 seconds (P < .001).

CONCLUSION

Massive aspiration of fluid occurs along the tracheal tube cuff during suction with the closed tracheal suction system.

Для цитирования: Dave Mital H, Frotzler Angela, Madjdpour Caveh, Koepfer Nelly, Weiss Markus. Massive aspiration past the tracheal tube cuff caused by closed tracheal suction system.. Journal of intensive care medicine. 2011. DOI: 10.1177/0885066610392516.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.