MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1038/s41390-026-05218-7

Медицинская библиотека · Pediatric research · 2026

Neonatal blood glucose monitoring using glucose estimation by phase delay between oxy- and deoxyhemoglobin.

Kusaka Tomoaki, Koyano Kosuke, Nakazawa Tomoya, Noguchi Yuta, Shinabe Yuta, Morita Hirosuke, Nishioka Katsufumi, Inoue Kota, Morimoto Aya, Nakamura Shinji, Kondo Sonoko, Konishi Yukihiko, Morishita Keiji, Kusaka Takashi

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Reliable, painless glucose surveillance is critical during the early neonatal fall in plasma glucose levels. However, existing continuous glucose monitors are invasive and expensive. We evaluated a non-invasive method for estimating blood glucose levels from the phase delay (Δθ) between oxy- and deoxyhemoglobin waveforms, expressed as a metabolic index (MI).

METHODS

Thirty-eight term neonates were enrolled on postnatal day 2, and 30 recordings satisfied the predefined signal quality criteria. Neonatal pulse oximetry probes captured 5-min photoplethysmography traces, from which stable 15-60-s epochs were extracted. The α-corrected MI for each infant was correlated with plasma glucose measured on an ABL90 FLEX analyzer. Glycemic slopes were compared with published adult data using ANCOVA.

RESULTS

The α-corrected MI showed a significant positive correlation with plasma glucose (R2 = 0.51, r = 0.71; p < 0.01). Linear regression for neonates was MI = 0.17 × glucose - 6.78. Both slope and intercept differed from adult values (p < 0.05), indicating age-dependent modulation of the Δθ-glucose relationship.

CONCLUSIONS

Phase-delay analysis using standard pulse oximetry provides a promising approach for continuous glucose estimation in term neonates. Age-specific calibration is required before clinical deployment; however, the technique's low cost and noninvasiveness potentiate universal bedside hypoglycemia surveillance.

IMPACT: Key message: Non-invasive glucose monitoring using the phase delay between oxy- and deoxyhemoglobin waveforms obtained from a pulse oximetry probe shows a significant correlation with blood glucose levels in term neonates. Literature contribution: This study demonstrates that the metabolic index derived from hemoglobin phase delays differs between neonates and adults, suggesting the need for age-specific calibration.

CLINICAL IMPACT: This low-cost technique may enable continuous bedside glucose surveillance in neonates without repeated blood sampling.

Для цитирования: Kusaka Tomoaki, Koyano Kosuke, Nakazawa Tomoya, Noguchi Yuta, Shinabe Yuta, Morita Hirosuke, Nishioka Katsufumi, Inoue Kota, Morimoto Aya, Nakamura Shinji, Kondo Sonoko, Konishi Yukihiko, Morishita Keiji, Kusaka Takashi. Neonatal blood glucose monitoring using glucose estimation by phase delay between oxy- and deoxyhemoglobin.. Pediatric research. 2026. DOI: 10.1038/s41390-026-05218-7.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.