Construction and characterization of an arterial input function measurement system for dynamic PET.

Carroll Lia, Daoud Youstina, Sarrhini Otman, Beaudoin Jean-François, Croteau Etienne, Convert Laurence, Lecomte Roger, Enger Shirin A

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Dynamic positron emission tomography is an under utilized clinical technique. It requires knowledge of the time-course activity concentration in the patient's arterial blood, called arterial input function (AIF), normally acquired via arterial blood sampling. Alternative methods to acquire the AIF do exist, but each has its own limitations. This study presents a non-invasive radiation detector, called Radian-PET, designed to measure the AIF non-invasively.

METHODS

10 cm long scintillating fibers arranged in two bands of 32 fibers were read out by a 64-channel silicon photomultiplier array. Calibration measurements were performed using 68Ge rod sources placed over the sensitive volume of the detector. Inter-channel variability was measured using the scans with the source perpendicular to the fibers. Two decay measurements were performed using 18F samples with varying activity. Cross-validation measurements were performed by placing a microfluidic blood sampling detector before Radian-PET in a microfluidic circuit. 18F, 11C and 68Ga injections were used to simulate the AIF.

RESULTS

Calibration measurements show that the electronic discriminator threshold, with minimal detectable activity below 100 kBq per mL for a 1-second integration window. The measured 18F half-lives were 109.75 ± 0.04 min and 110.82 ± 0.73 min and agreed within uncertainties with the literature. Linear regressions from the cross-validation measurements showed good agreement (R2> 0.96) for all scans.

CONCLUSIONS

Radian-PET can accurately measure clinically relevant activity concentrations and detect quickly changing activity levels like those used to perform kinetic analysis. Further work is required to increase detector efficiency and reduce sources of noise.

Для цитирования: Carroll Lia, Daoud Youstina, Sarrhini Otman, Beaudoin Jean-François, Croteau Etienne, Convert Laurence, Lecomte Roger, Enger Shirin A. Construction and characterization of an arterial input function measurement system for dynamic PET.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2026. DOI: 10.1016/j.ejmp.2026.105804.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.