DOI 10.1099/jmm.0.001881

Медицинская библиотека · Journal of medical microbiology · 2024

Evaluating the accuracy of the MBT Lipid Xtract Kit for assessing colistin resistance in comparison to broth microdilution.

Lo Calvin Ka-Fung, Ritchie Gordon, Bilawka Jennifer, Gowland Leah, Chorlton Samuel D, Jang Willson, Matic Nancy, Romney Marc G, Stefanovic Aleksandra, Lowe Christopher F

СтатьяСравнительное исследованиеОценочное исследованиеПеревод пока не подготовлен

Аннотация

Colistin resistance testing methods such as broth microdilution (BMD) are time-consuming and labour intensive for clinical laboratories. MBT Lipid Xtract Kit on MALDI Biotyper Sirius System (Bruker, Billerica, MA, USA) utilizes lipidomic analysis to identify specific cell wall modifications associated with colistin resistance. We compared MBT to BMD (ComASP Colistin, Liofilchem) across 36 Gram-negative isolates (non-resistant MIC ≤2 µg ml-1, resistant MIC ≥4 µg ml-1). All samples were tested twice on MBT with discrepant results repeated before assessing categorical agreement between MBT and BMD. 44.4% (16/36) of isolates were colistin resistant via BMD. MBT Lipid Xtract had 80.6% agreement (29/36) with BMD, with 5/7 discrepancies corrected to match upon repeat testing. There was 100% agreement for Escherichia coli isolates (n=16). The whole-genome sequencing was completed on the two discrepant Klebsiella pneumoniae isolates, with variants within colistin resistance-associated loci identified (MIC 0.5 µg ml-1: arnC S30T, pmrB T246A, lapB N212T, lpxM S253G, crrB Q287K and MIC >16 µg ml-1: arnC S30T, pmrB R90insRN, pmrB T246A, pmrA E57G, lpxM S253G). Further evaluation, particularly for non-E. coli, of MBT is required prior to implementation in clinical laboratories.

Для цитирования: Lo Calvin Ka-Fung, Ritchie Gordon, Bilawka Jennifer, Gowland Leah, Chorlton Samuel D, Jang Willson, Matic Nancy, Romney Marc G, Stefanovic Aleksandra, Lowe Christopher F. Evaluating the accuracy of the MBT Lipid Xtract Kit for assessing colistin resistance in comparison to broth microdilution.. Journal of medical microbiology. 2024. DOI: 10.1099/jmm.0.001881.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.