MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.70530

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026

Computed Tomography Derived Procedure Simulations for Redo Transcatheter Aortic Valve Replacement.

van den Dorpel Mark M P, de Backer Ole, Abdel-Wahab Mohamed, Beneduce Alessandro, Dumonteil Nicolas, Blackman Daniel J, Willemen Yannick, Dargan James, Poon Karl, Daemen Joost, de Assis Lucas Uchoa, van Niekerk Jenna, Kardys Isabella, Tchetche Didier, De Beule Matthieu, Jan Nuis Rutger, Rocatello Giorgia, Debusschere Nic, Mortier Peter, Van Mieghem Nicolas M

СтатьяНаблюдательное исследованиеМногоцентровое исследованиеВалидационное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Transcatheter aortic valve (TAV) failure may require redo transcatheter aortic valve replacement (redo-TAVR). The multiple layers of metal after redo-TAVR may jeopardize coronary artery patency and accessibility and result in prosthesis-patient mismatch. Dedicated software may use multi-slice computed tomography (MSCT) imaging to simulate TAV deployment relative to the prior TAV and the individual's anatomy and predict coronary accessibility and TAV expansion.

AIM: We aimed to validate redo-TAVR simulations created in FEops HEARTguide.

METHODS

This multicenter observational study included redo-TAVR patients who had MSCT imaging available at baseline, after index-TAVR and after redo-TAVR. Patient-specific redo-TAVR simulations were created. Simulated valve-to-aorta distances (a proxy for coronary accessibility), neoskirt height, TAV expansion at multiple frame levels, and residual valve area after redo-TAVR were compared to follow-up MSCT.

RESULTS

Among 25 patients, HEARTguide simulations of valve-to-coronary (VTC) distance (left VTC 6.0 ± 2.2 vs. 6.4 ± 2.3 mm, ICC = 0.856, p < 0.001; right VTC 5.8 ± 1.9 vs. 5.6 ± 1.9 mm, ICC = 0.740, p < 0.001) showed good agreement while valve-to-sinotubular-junction (VTSTJ) distance showed moderate agreement with follow-up MSCT (left VTSTJ 2.5 ± 1.8 vs. 3.3 ± 1.8 mm, ICC = 0.614, p = 0.004; right VTSTJ 2.8 ± 1.8 vs. 2.9 ± 1.2 mm, ICC = 0.613, p = 0.003). Redo-TAVR frame expansion simulations (mean frame area 434.8 ± 141.6 vs. 401.6 ± 142.9 mm, ICC = 0.930, p < 0.001; 393.5 ± 63.2 vs. 383.1 ± 82.5 mm, ICC = 0.822, p < 0.001, respectively) and simulations of residual valve area after redo-TAVR showed good agreement for self-expanding and balloon-expandable TAV (407.8 ± 67.8 vs. 389.6 ± 50.4 mm, ICC = 0.837, p < 0.001; 358.0 ± 66.1 vs. 336.1 ± 79.9 mm, ICC = 0.823, p < 0.001).

CONCLUSIONS

In this pilot study, MSCT-derived simulations of redo-TAVR provided accurate insights into coronary accessibility, TAV expansion, and residual valve area.

Для цитирования: van den Dorpel Mark M P, de Backer Ole, Abdel-Wahab Mohamed, Beneduce Alessandro, Dumonteil Nicolas, Blackman Daniel J, Willemen Yannick, Dargan James, Poon Karl, Daemen Joost, de Assis Lucas Uchoa, van Niekerk Jenna, Kardys Isabella, Tchetche Didier, De Beule Matthieu, Jan Nuis Rutger, Rocatello Giorgia, Debusschere Nic, Mortier Peter, Van Mieghem Nicolas M. Computed Tomography Derived Procedure Simulations for Redo Transcatheter Aortic Valve Replacement.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2026. DOI: 10.1002/ccd.70530.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.