The Verification of Co-Axiality of the Three Self-Expanding Transcatheter Aortic Valve Systems According to the Difference in Shaft Spine.

Teng Yun, Enta Yusuke, Saigan Makoto, Satomi Natsuko, Kobayashi Yuta, Nakashima Masaki, Kosuga Momo, Maeda Manabu, Toki Yusuke, Munehisa Yoshiko, Miyasaka Masaki, Taguri Masataka, Toyoda Shigeru, Tada Norio

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Accurate bioprosthesis implantation depth during transcatheter aortic valve replacement (TAVR) is crucial for optimal outcomes. Self-expanding valves (SEVs) have evolved delivery catheter designs, from a double spine shaft in Evolut PRO+ (PRO + ) to single spine in Evolut FX (FX) and non-spine in Navitor, to improve access deliverability.

AIMS: To investigate the influence of different spine structures on valve implantation.

METHODS

Retrospectively, we investigated 95 severe aortic stenosis patients who underwent transfemoral TAVR using SEVs (PRO + : 41, FX: 21, and Navitor: 33 patients) from January 2021 to July 2023. The implantation depth difference (IDD) between Left Coronary Cusp (LCC) and Non-Coronary Cusp (NCC) was measured angiographically at the point of no recapture, defined as subtracting NCC implantation depth from LCC. Positive IDD indicates deeper LCC than NCC, also shown as "vertical axis." Whereas negative IDD indicates shallower LCC than NCC, demonstrated as "horizontal axis."

RESULTS

IDD at the point of no recapture showed significant difference among three groups (PRO + : 2.42 [0.81-3.42], FX: 1.01 [0.29-1.77], Navitor: -2.16[-2.74- (-1.14)] mm, p < 0.001, PRO + > FX/Navitor, FX > Navitor). PRO+ had a positive IDD, indicating deeper LCC implantation. Navitor had a negative IDD, demonstrating shallower implantation on LCC, and FX showed co-axial implantation.

CONCLUSION

PRO+ with double spines showed vertical axis, Navitor with non-spine showed horizontal axis, whereas FX demonstrated co-axiality at the point of no recapture. This knowledge may help refine the manipulation of procedural implantation based on the shaft rigidity in each device to gain co-axiality.

Для цитирования: Teng Yun, Enta Yusuke, Saigan Makoto, Satomi Natsuko, Kobayashi Yuta, Nakashima Masaki, Kosuga Momo, Maeda Manabu, Toki Yusuke, Munehisa Yoshiko, Miyasaka Masaki, Taguri Masataka, Toyoda Shigeru, Tada Norio. The Verification of Co-Axiality of the Three Self-Expanding Transcatheter Aortic Valve Systems According to the Difference in Shaft Spine.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2026. DOI: 10.1002/ccd.70512.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.