DOI 10.3109/00016489.2010.507779

Медицинская библиотека · Acta oto-laryngologica · 2010

Intraoperative adjustments to optimize active middle ear implant performance.

Tringali Stéphane, Koka Kanthaiah, Deveze Arnaud, Ferber Alexander T, Jenkins Herman A, Tollin Daniel J

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

CONCLUSION

After initial contact of the active middle ear implant (AMEI) on the incus, significant increases in device performance can be achieved intraoperatively without affecting residual hearing by additional static loading of the incus with 62 μm (quarter turn) to 125 μm (half turn) increments via an adjustment screw.

OBJECTIVES

To assess the performance gains of driving the incus with an AMEI under increasing static loads in cadaveric temporal bones.

METHODS

Incus drive efficacy was assessed using laser Doppler velocimetry measurements of stapes velocities over a frequency range of 0.25 to 8 kHz. Results were compared to stapes velocities following acoustic stimulation via insert earphone. Maximum equivalent ear canal sound pressure level (L(Emax)) and residual hearing loss after initial loading of the AMEI (first contact) were compared in each temporal bone. Additional increases in incus load were induced by turning an adjustment screw in quarter turn steps, corresponding to 62 μm increments per step. L(Emax)and residual hearing loss were reassessed after each step. For each temporal bone, experiments were repeated for three different AMEIs.

RESULTS

On average across bones, incus stimulation upon initial contact produced an L(Emax)of 125, 127, and 121 dB SPL and residual hearing losses of -2, -1, and -1 dB with respect to unloaded, unaided conditions for the three AMEIs, respectively. Across bones and transducers, increasing static transducer load by incrementing the AMEI up to 125 μm significantly improved performance without affecting residual hearing loss. Loading beyond 125 μm (half turn) did not improve performance but significantly increased residual hearing loss.

Для цитирования: Tringali Stéphane, Koka Kanthaiah, Deveze Arnaud, Ferber Alexander T, Jenkins Herman A, Tollin Daniel J. Intraoperative adjustments to optimize active middle ear implant performance.. Acta oto-laryngologica. 2010. DOI: 10.3109/00016489.2010.507779.