Excised larynx evaluation of wedge-shaped adjustable balloon implant for minimally invasive type I thyroplasty.

СтатьяПоддержка NIH: внешняя программаНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Hoffman Matthew R, Devine Erin E, McCulloch Timothy M, Jiang Jack J

Аннотация

OBJECTIVES/HYPOTHESIS: To describe the method of inserting a wedge-shaped adjustable balloon implant (wABI) via a minithyrotomy for medialization thyroplasty and evaluate its effect on a range of phonatory parameters using the excised larynx bench apparatus.

STUDY DESIGN: Repeated measures with each larynx serving as its own control.

METHODS: A prototype wABI was deployed in six excised canine larynges of various sizes through a minithyrotomy and then filled with saline. Mucosal wave, aerodynamic, and acoustic parameters were measured for three conditions: normal, vocal fold paralysis, and paralysis with the wABI.

RESULTS: Phonation threshold pressure (P < .001), flow (P < .001), and power (P = .002) were significantly lower for wABI compared to paralysis trials; values did not differ significantly from normal trials. Percent jitter (P = .002) and percent shimmer (P = .007) were also significantly decreased compared to the paralysis condition, and values were not significantly different compared to normal. The mucosal wave was preserved after insertion of the wABI.

CONCLUSIONS: Effective vocal fold medialization with preservation of the mucosal wave was observed with the wABI in this preliminary excised larynx experiment. The wABI offers the potential for a minimally invasive insertion in addition to postoperative adjustability. Further studies in living animals and humans are warranted to evaluate clinical utility.

LEVEL OF EVIDENCE: NA.