MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/lary.21431

Медицинская библиотека · The Laryngoscope · 2011

Preliminary investigation of adjustable balloon implant for type I thyroplasty.

Hoffman Matthew R, Witt Rachel E, McCulloch Timothy M, Jiang Jack J

СтатьяПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

OBJECTIVE

We present the adjustable balloon implant (ABI), a novel implant to be used in type I thyroplasty for the treatment of vocal fold paralysis. The ABI offers the same medialization provided by other implants, but can easily be catered to individual patient anatomy as well as modified postoperatively without the need for a revision thyroplasty.

STUDY

DESIGN

Repeated measures with each larynx serving as its own control.

METHODS

Medialization thyroplasty (MT) with the ABI was performed on five excised canine larynges. Mucosal wave, aerodynamic, and acoustic parameters were measured for three conditions: normal; right vocal fold paralysis; and paralysis with the ABI.

RESULTS

Insertion of the ABI resulted in significant decreases in both phonation threshold pressure and phonation threshold flow. Perturbation parameters of percent jitter and percent shimmer were also significantly decreased and restored to normal levels. Signal-to-noise ratio was significantly increased to the normal level as well. The mucosal wave was preserved after implant insertion.

CONCLUSIONS

This preliminary experiment showing significant improvements in aerodynamic and acoustic parameters demonstrates the potential of the ABI as a thyroplasty implant. Effective medialization and preservation of the mucosal wave combined with postoperative adjustability makes it a potentially valuable clinical device.

Для цитирования: Hoffman Matthew R, Witt Rachel E, McCulloch Timothy M, Jiang Jack J. Preliminary investigation of adjustable balloon implant for type I thyroplasty.. The Laryngoscope. 2011. DOI: 10.1002/lary.21431.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.