Comparison of the Accuracy of Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy in Localizing Rectus Muscle Insertions.

Mirmohammadsadeghi Arash, Hamzeh Nikoo, Ghassemi Fariba, Akbari Mohammad Reza

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the accuracy of anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM) in localizing rectus muscle insertions.

METHODS

The study was performed on 27 patients (39 rectus muscles) who required primary or secondary surgery. Using caliper function in the AS-OCT and UBM software, the distance from the insertion site to the anterior chamber angle was measured. The actual muscle insertion distance from limbus was considered as the measured distance plus 1 mm. The measurements by UBM and AS-OCT were compared with intraoperative measurements and with each other.

RESULTS

AS-OCT and UBM were performed on 13 medial rectus, 24 lateral rectus, and 2 superior rectus muscles. Ninety two percent of UBM measurements (36 muscles) were within 1 mm, one was within 1-1.5 mm, and 2 were within 1.5-2 mm of surgery measurements. Eighty five percent of AS-OCT measurements (33 muscles) were within 1 mm, 5 were within 1-1.5 mm, and one was within 1.5-2 mm of surgery measurements. In all cases, the mean absolute error of the UBM (0.54 ± 0.44) and AS-OCT (0.51 ± 0.36) showed no significant difference (p = .76).

CONCLUSION

AS-OCT and UBM can be used interchangeably to localize rectus muscle insertions and showed good agreement with intraoperative measurements.

Для цитирования: Mirmohammadsadeghi Arash, Hamzeh Nikoo, Ghassemi Fariba, Akbari Mohammad Reza. Comparison of the Accuracy of Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy in Localizing Rectus Muscle Insertions.. Journal of binocular vision and ocular motility. 2022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.