Nihon Hoshasen Gijutsu Gakkai zasshi · 2022
[Three-dimensional Quantitative Evaluation Method in 123I-MIBG Myocardial SPECT-CT].
Аннотация на английскомСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yasumoto Yoshiaki, Daisaki Hiromitsu, Nakahara Tadaki, Ito Ryoichi, Fujita Isao
Аннотация
PURPOSE: To distinguish neurodegenerative diseases using 123I-metaiodobenzylguanidine (MIBG). This study proposes a method to evaluate myocardial standardized uptake value (SUV) and assess its accuracy.
METHODS: We created a 17-segment polar map of the myocardial region from single-photon emission computed tomography-computed tomography (SPECT-CT) images using a cardioliver phantom simulating the standard uptake of MIBG. We clarified the optimal reconstruction conditions with good repeatability and accuracy of quantitative values and compared them with the H/M ratio. Myocardial SUVs were evaluated from eight normal cases using our method established from the phantom experiment and compared with the H/M ratio.
RESULTS: The optimal numbers of iterations and subsets in OSEM reconstruction were both 10. The optimal full width at half maximum (FWHM) value of the Gaussian filter was 4 pixels. The RCs and %CV of (1) maximum SUVmax (MaxSUVmax) and (2) average SUVmax (AveSUVmax) were (1) 36.5% and 4.99%, and (2) 33.6% and 4.84%, respectively. The RC and %CV of the H/M ratio was 15.0% and 1.50%, respectively. In clinical cases, average MaxSUVmax and AveSUVmax were 8.27 and 7.58, respectively.
CONCLUSION: Myocardial SUV can provide quantitative values slightly closer to theoretical values than the H/M ratios. Besides, using the optimal reconstruction parameters makes it feasible to quantitatively assess myocardial uptake with good repeatability.
METHODS: We created a 17-segment polar map of the myocardial region from single-photon emission computed tomography-computed tomography (SPECT-CT) images using a cardioliver phantom simulating the standard uptake of MIBG. We clarified the optimal reconstruction conditions with good repeatability and accuracy of quantitative values and compared them with the H/M ratio. Myocardial SUVs were evaluated from eight normal cases using our method established from the phantom experiment and compared with the H/M ratio.
RESULTS: The optimal numbers of iterations and subsets in OSEM reconstruction were both 10. The optimal full width at half maximum (FWHM) value of the Gaussian filter was 4 pixels. The RCs and %CV of (1) maximum SUVmax (MaxSUVmax) and (2) average SUVmax (AveSUVmax) were (1) 36.5% and 4.99%, and (2) 33.6% and 4.84%, respectively. The RC and %CV of the H/M ratio was 15.0% and 1.50%, respectively. In clinical cases, average MaxSUVmax and AveSUVmax were 8.27 and 7.58, respectively.
CONCLUSION: Myocardial SUV can provide quantitative values slightly closer to theoretical values than the H/M ratios. Besides, using the optimal reconstruction parameters makes it feasible to quantitatively assess myocardial uptake with good repeatability.