Replacement of fluoroscopy by ultrasonography in the evaluation of hemidiaphragm function, an exploratory prospective study.

Skaarup Søren Helbo, Juhl-Olsen Peter, Grundahl Anne Sofie, Løgstrup Brian Bridal

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Dysfunction of the diaphragm may ultimately lead to respiratory insufficiency and compromise patient outcome. Evaluation of diaphragm function is cumbersome. Fluoroscopy has been the gold standard to measure diaphragmatic excursion. Ultrasonography can visualize diaphragm excursion and holds many advantages such as no radiation exposure, increased portability and accessibility. However, correlation between fluoroscopy and ultrasonography has never been studied. We aimed to compare fluoroscopic and ultrasound measures of diaphragm excursion to determine if ultrasonography can replace fluoroscopy.

METHODS

We performed ultrasound and fluoroscopy simultaneously during sniff inspiration and at total inspiratory capacity in patients with chronic obstructive pulmonary disease, heart failure and in healthy volunteers. Cranio-caudal excursion was measured by fluoroscopy and compared directly to M-mode excursion, B-mode excursion, area change, resting thickness, thickening fraction and contraction velocity measured by ultrasonography.

RESULTS

Forty-two participants were included. The Pearson correlation between M-mode and fluoroscopy excursion was 0.61. The slope was 0.9 (90%CI 0.76-1.04) in a regression analysis. Using the Bland-Altman method, the bias was - 0.39 cm (95% CI - 1.04-0.26), p = 0.24. The Pearson correlation between fluoroscopy and B-mode and area change ultrasonography was high; low for thickness and fraction. All correlations were lower during sniff inspiration compared with inspiratory capacity breathing.

CONCLUSION

Ultrasonography has an acceptable correlation and bias compared to fluoroscopy and can thus be used as the primary tool to evaluate diaphragm excursion.

Для цитирования: Skaarup Søren Helbo, Juhl-Olsen Peter, Grundahl Anne Sofie, Løgstrup Brian Bridal. Replacement of fluoroscopy by ultrasonography in the evaluation of hemidiaphragm function, an exploratory prospective study.. The ultrasound journal. 2024. DOI: 10.1186/s13089-023-00355-0.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.