Role of the oxygen supply position in the circuit during continuous positive airway pressure therapy: a randomized cross-over simulation study.

Bousraou Zoe, Benning Lara, Spengler Christina, Ulrich Silvia, Lüchinger Martin, Schwarz Esther Irene

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Continuous Positive Airway Pressure (CPAP) is the main treatment for patients with obstructive sleep apnoea (OSA). Hypoxemic patients with sleep-disordered breathing may require CPAP and additional low-flow oxygen. However, specific recommendations on where to attach the oxygen bleed-in connector are lacking.

METHODS

This randomised cross-over trial assessed the fraction of inspired oxygen (FIO₂) in a simulated OSA patient’s lung resulting from different rates of low-flow oxygen (1, 2 and 3 l/min) added via three different bleed-in connector locations under different levels of either fixed pressure (5, 9 and 12 cmH2O) or constant flow (5, 10 and 15 l/min) (repeated 3x in random order). The three locations tested in a single-limb circuit with a vented full-face mask were: (1) near the CPAP mask between the mask and tubing, (2) at the tube outlet of the device, and (3) directly connected to the oxygen-port of the device.

RESULTS

The bleed-in connector positioned near the tube outlet generally produced the highest and near the CPAP mask the lowest FiO₂. At the highest pressure and flow, FiO₂ remained more stable across the locations and oxygen flows of 1–3 l/min, compared to lower pressure and lower flow. FiO₂ differences between locations were notably more pronounced under different flows, ranging from 0 to 22% relative change, compared to a more limited range of 0 to 8.3% under varying CPAP pressures.

CONCLUSION

We recommend positioning the oxygen bleed-in connector at the device tube outlet for CPAP devices without an oxygen port, particularly when using lower pressures.

Для цитирования: Bousraou Zoe, Benning Lara, Spengler Christina, Ulrich Silvia, Lüchinger Martin, Schwarz Esther Irene. Role of the oxygen supply position in the circuit during continuous positive airway pressure therapy: a randomized cross-over simulation study.. Sleep & breathing = Schlaf & Atmung. 2026. DOI: 10.1007/s11325-026-03589-z.

Предметные рубрики MeSH