International ophthalmology · 2025

Comparing toric intraocular lens alignment: intraoperative image-guided system versus manual marking in cataract surgery: a randomized clinical trial.

СтатьяРандомизированное исследованиеСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Tripathi Amit Nandan, Kumar Santosh, Sharma Vijay Kumar, Rana Vipin, Kakati Kalpamoi, Goenka Ranjit, Bandopadhyay Sandepan

Аннотация

PURPOSE: To compare the accuracy of toric intraocular lens (IOL) alignment using the image-guided Callisto Eye system (M/S Carl Zeiss Meditech AG, Germany) versus the conventional manual marking method (Bubble marker) in cataract surgery.

DESIGN: Prospective, single-centre, double-masked, randomized controlled trial with two comparative arms. The study was registered with the Clinical Trial Registry of India (CTRI/2023/07/055833).

METHODS: A total of 76 patients aged 18 years or older, with cataract and corneal astigatism greater than 1.25 diopters (D), were enrolled. Patients were randomly assigned to either the image-guided group (Group 1 n = 38) or the manual marking group (Group 2 n = 38). Group 1 underwent IOL alignment using the Callisto Eye system, while Group 2 used the bubble marker manual technique. The primary outcome measure was IOL misalignment on postoperative day (POD) 1. Secondary outcomes included rotational stability, residual astigmatism, and uncorrected distance visual acuity (UDVA) on POD 30.

RESULTS: The mean IOL misalignment on POD 1 was 2.7 ± 3.6 degrees in Group 1 and 4 ± 3.1 degrees in Group 2, with no statistically significant difference (P = 0.108). Rotational stability showed a mean change of 0.71 ± 1.3 degrees in Group 1 and 0.78 ± 1.5 degrees in Group 2 (P = 0.815). Residual astigmatism was comparable between the groups (P = 0.176), and UDVA was similar, with no significant differences (P = 0.595).

CONCLUSION: Both the image-guided Callisto Eye system and the manual marking method demonstrated statistically equivalent effectiveness in toric IOL alignment, rotational stability, residual astigmatism, and UDVA.