Igarashi Yutaka, Ikeda Shimpei, Hirai Kunio, Tominaga Naoki, Mizobuchi Taiki, Shigeta Kenta, Ishii Hiromoto, Yokobori Shoji
Аннотация
BACKGROUND: Many invasive procedures performed in the emergency room (ER) have potential risks and complications. Because of their limitations, in particular size, portable X-ray devices are generally not used during such procedures. However, miniaturized units enable physicians to obtain X-ray images without assistance.
METHODS: We developed a safe, compact, and lightweight X-ray unit that was used during five invasive procedures in the ER. In all procedures, a chest X-ray image was obtained, to confirm the utility of the unit.
RESULTS: Case 1 (central venous catheter placement): After needle and guidewire insertion and catheter placement, the location of the catheter could be confirmed. Case 2 (chest tube insertion): During insertion of a chest tube into the pleural space, the tip of the thoracic tube was confirmed to be at the correct location. Case 3 (percutaneous tracheostomy or cricothyroidotomy): After needle and guidewire insertion, the guidewire was visualized in the right main bronchus and was inserted into the trachea. Case 4 (resuscitative endovascular aortic balloon of the aorta): The captured image revealed that the catheter was located in zone I before balloon inflation. Case 5 (Sengstaken-Blakemore tube): The image revealed that the balloon was located in the stomach.
CONCLUSIONS: The portable X-ray unit improved medical safety during invasive procedures frequently performed in the ER.
METHODS: We developed a safe, compact, and lightweight X-ray unit that was used during five invasive procedures in the ER. In all procedures, a chest X-ray image was obtained, to confirm the utility of the unit.
RESULTS: Case 1 (central venous catheter placement): After needle and guidewire insertion and catheter placement, the location of the catheter could be confirmed. Case 2 (chest tube insertion): During insertion of a chest tube into the pleural space, the tip of the thoracic tube was confirmed to be at the correct location. Case 3 (percutaneous tracheostomy or cricothyroidotomy): After needle and guidewire insertion, the guidewire was visualized in the right main bronchus and was inserted into the trachea. Case 4 (resuscitative endovascular aortic balloon of the aorta): The captured image revealed that the catheter was located in zone I before balloon inflation. Case 5 (Sengstaken-Blakemore tube): The image revealed that the balloon was located in the stomach.
CONCLUSIONS: The portable X-ray unit improved medical safety during invasive procedures frequently performed in the ER.