Evaluation of a semi-automated von Willebrand factor multimer assay, the Hydragel 5 von Willebrand multimer, by two European Centers.

Bowyer Annette E, Goodfellow Karen J, Seidel Holger, Westhofen Philipp, Stufano Francesca, Goodeve Anne, Kitchen Stephen, Makris Michael

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The phenotypic diagnosis of von Willebrand disease (VWD) is a multistep process with classification dependent on the quantification of von Willebrand factor (VWF) multimeric structure. VWF multimer analysis is a technically challenging, lengthy and non-standardised assay, usually performed in specialist laboratories. Recently, a new semi-automated multimer assay, the Hydragel 5 von Willebrand multimers (H5VWM) has become available.

OBJECTIVES

This study, performed in two European centres, compared existing in-house multimer assays to the H5VWM in individuals with and without VWD.

RESULTS

Overall agreement of 91.1% was observed in 74 individuals with normal VWF levels, 57 patients grouped as type 1 VWD, 33 type 2A, 16 type 2B, 28 type 2M, 11 type 2N. Patients tested following Desmopressin or VWF concentrate, with thrombotic thrombocytopenic purpura and acquired von Willebrand syndrome were also evaluated. Many of the discrepancies between methods were in patients with genetic mutations linked to more than one type of VWD including p.R1374C/H and p.R1315C. Quantifiable multimer results were available within one working day. Densitometry improved the interpretation of the multimers with slight structural variations that were not apparent by visual inspection of the in-house method.

CONCLUSIONS

5VWM was a rapid, sensitive, standardised assay which used existing technology and could be included as an initial screen of VWF multimers in a VWD diagnostic algorithm in conjunction with traditional multimer analysis.

Для цитирования: Bowyer Annette E, Goodfellow Karen J, Seidel Holger, Westhofen Philipp, Stufano Francesca, Goodeve Anne, Kitchen Stephen, Makris Michael. Evaluation of a semi-automated von Willebrand factor multimer assay, the Hydragel 5 von Willebrand multimer, by two European Centers.. Research and practice in thrombosis and haemostasis. 2018. DOI: 10.1002/rth2.12141.